Factors Affecting Mydriasis-Free Flicker ERGs Recorded With Real-Time Correction for Retinal Illuminance: Study of 150 Young Healthy Subjects
- PMID: 29049731
- DOI: 10.1167/iovs.17-22587
Factors Affecting Mydriasis-Free Flicker ERGs Recorded With Real-Time Correction for Retinal Illuminance: Study of 150 Young Healthy Subjects
Abstract
Purpose: A small, full-field flicker electroretinogram (ERG) recording system was recently developed to record flicker ERGs without mydriasis (RETeval). The device delivers a stimulus with constant retinal illuminance by adjusting the retinal luminance to compensate for changes in the pupillary area. The purpose of this study was to determine what factors affect the fundamental components of the flicker ERGs recorded by RETeval in young healthy subjects.
Methods: Flicker ERGs were recorded with the RETeval system from 150 eyes of 150 young healthy subjects (age, 20-29 years). Univariate and multivariate linear regression analyses were performed to identify the factors that affected the implicit times and amplitudes of the fundamental component of the flicker ERGs. The independent variables included age, sex, refractive error, axial length, and pupillary area.
Results: Multivariate regression analyses indicated that a longer axial length (P = 0.03) and larger pupillary area (P = 0.008) were independent factors that were significantly associated with longer implicit times of the fundamental component of the flicker ERGs. Multivariate regression analyses also showed that the female sex (P = 0.03) was an independent factor, which was significantly associated with larger amplitude fundamental component of the flicker ERGs.
Conclusions: These results indicate that the fundamental components of the RETeval flicker ERGs are significantly affected by the axial length, pupillary area, and sex of young healthy subjects. The results also suggest that it would be better to compensate for the Stiles-Crawford effect when flicker ERGs are recorded with natural pupils.
Similar articles
-
Assessment of factors affecting flicker ERGs recorded with RETeval from data obtained from health checkup screening.PLoS One. 2023 Apr 24;18(4):e0284686. doi: 10.1371/journal.pone.0284686. eCollection 2023. PLoS One. 2023. PMID: 37093860 Free PMC article.
-
Effect of Pupil Size on Flicker ERGs Recorded With RETeval System: New Mydriasis-Free Full-Field ERG System.Invest Ophthalmol Vis Sci. 2015 Jun;56(6):3684-90. doi: 10.1167/iovs.14-16349. Invest Ophthalmol Vis Sci. 2015. PMID: 26047169
-
Effects of recording sequence on flicker electroretinographics recorded with natural pupils corrected for pupil area.Acta Ophthalmol. 2021 Jun;99(4):411-417. doi: 10.1111/aos.14618. Epub 2020 Sep 10. Acta Ophthalmol. 2021. PMID: 32914561 Free PMC article.
-
Constant luminance (cd·s/m2) versus constant retinal illuminance (Td·s) stimulation in flicker ERGs.Doc Ophthalmol. 2017 Apr;134(2):75-87. doi: 10.1007/s10633-017-9572-3. Epub 2017 Feb 3. Doc Ophthalmol. 2017. PMID: 28160194 Free PMC article.
-
Flicker electroretinograms recorded with mydriasis-free RETeval system before and after cataract surgery.Eye (Lond). 2017 Nov;31(11):1589-1593. doi: 10.1038/eye.2017.110. Epub 2017 Jun 16. Eye (Lond). 2017. PMID: 28622319 Free PMC article.
Cited by
-
Paediatric norms for photopic electroretinogram testing based on a large cohort of Chinese preschool children.BMJ Open Ophthalmol. 2024 Feb 21;9(1):e001393. doi: 10.1136/bmjophth-2023-001393. BMJ Open Ophthalmol. 2024. PMID: 38388004 Free PMC article.
-
Assessment of factors affecting flicker ERGs recorded with RETeval from data obtained from health checkup screening.PLoS One. 2023 Apr 24;18(4):e0284686. doi: 10.1371/journal.pone.0284686. eCollection 2023. PLoS One. 2023. PMID: 37093860 Free PMC article.
-
Transient Increase of Flicker Electroretinography Amplitudes after Cataract Surgery: Association with Postoperative Inflammation.Ophthalmol Sci. 2022 Nov 4;3(1):100243. doi: 10.1016/j.xops.2022.100243. eCollection 2023 Mar. Ophthalmol Sci. 2022. PMID: 36545261 Free PMC article.
-
Comparing the RETeval® portable ERG device with more traditional tabletop ERG systems in normal subjects and selected retinopathies.Doc Ophthalmol. 2023 Apr;146(2):137-150. doi: 10.1007/s10633-022-09903-w. Epub 2022 Oct 23. Doc Ophthalmol. 2023. PMID: 36273363 Free PMC article.
-
Novel Machine-Learning Based Framework Using Electroretinography Data for the Detection of Early-Stage Glaucoma.Front Neurosci. 2022 May 4;16:869137. doi: 10.3389/fnins.2022.869137. eCollection 2022. Front Neurosci. 2022. PMID: 35600610 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources