Videolaryngoscopy for Physician-Based, Prehospital Emergency Intubation: A Prospective, Randomized, Multicenter Comparison of Different Blade Types Using A.P. Advance, C-MAC System, and KingVision

Anesth Analg. 2018 May;126(5):1565-1574. doi: 10.1213/ANE.0000000000002735.

Abstract

Background: Videolaryngoscopy is a valuable technique for endotracheal intubation. When used in the perioperative period, different videolaryngoscopes vary both in terms of technical use and intubation success rates. However, in the prehospital environment, the relative performance of different videolaryngoscopic systems is less well studied.

Methods: We conducted this prospective, randomized, multicenter study at 4 German prehospital emergency medicine centers. One hundred sixty-eight adult patients requiring prehospital emergency intubation were treated by an emergency physician and randomized to 1 of 3 portable videolaryngoscopes (A.P. Advance, C-MAC PM, and channeled blade KingVision) with different blade types. The primary outcome variable was overall intubation success and secondary outcomes included first-attempt intubation success, glottis visualization, and difficulty with handling the devices. P values for pairwise comparisons are corrected by the Bonferroni method for 3 tests (P[BF]). All presented P values are adjusted for center.

Results: Glottis visualization was comparable with all 3 devices. Overall intubation success for A.P. Advance, C-MAC, and KingVision was 96%, 97%, and 61%, respectively (overall: P < .001, A.P. Advance versus C-MAC: odds ratio [OR], 0.97, 95% confidence interval [CI], 0.13-7.42, P[BF] > 0.99; A.P. Advance versus KingVision: OR, 0.043, 95% CI, 0.0088-0.21, P[BF] < 0.001; C-MAC versus KingVision: OR, 0.043, 95% CI, 0.0088-0.21, P[BF] < 0.001). Intubation success on the first attempt with A.P. Advance, C-MAC, and KingVision was 86%, 85%, and 48%, respectively (overall: P < .001, A.P. Advance versus C-MAC: OR, 0.89, 95% CI, 0.31-2.53, P[BF] > 0.99; A.P. Advance versus KingVision: OR, 0.24, 95% CI, 0.055-0.38, P[BF] = 0.0054; C-MAC versus KingVision: OR, 0.21, 95% CI, 0.043-.34, P[BF] < 0.003). Direct laryngoscopy for successful intubation with the videolaryngoscopic device was necessary with the A.P. Advance in 5 patients, and with the C-MAC in 4 patients. In the KingVision group, 21 patients were intubated with an alternative device.

Conclusions: During prehospital emergency endotracheal intubation performed by emergency physicians, success rates of 3 commercially available videolaryngoscopes A.P. Advance, C-MAC PM, and KingVision varied markedly. We also found that although any of the videolaryngoscopes provided an adequate view, actual intubation was more difficult with the channeled blade KingVision.

Trial registration: ClinicalTrials.gov NCT01635660.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Airway Management / instrumentation
  • Airway Management / methods*
  • Airway Management / standards
  • Emergency Medical Services / methods*
  • Emergency Medical Services / standards
  • Female
  • Humans
  • Intubation, Intratracheal / instrumentation
  • Intubation, Intratracheal / methods*
  • Intubation, Intratracheal / standards
  • Laryngoscopy / instrumentation
  • Laryngoscopy / methods*
  • Laryngoscopy / standards
  • Male
  • Middle Aged
  • Physician's Role*
  • Prospective Studies
  • Video Recording / methods*
  • Video Recording / standards
  • Young Adult

Associated data

  • ClinicalTrials.gov/NCT01635660