Estimating Energy Expenditure with ActiGraph GT9X Inertial Measurement Unit
- PMID: 29271847
- DOI: 10.1249/MSS.0000000000001532
Estimating Energy Expenditure with ActiGraph GT9X Inertial Measurement Unit
Abstract
Purpose: The purpose of this study was to explore whether gyroscope and magnetometer data from the ActiGraph GT9X improved accelerometer-based predictions of energy expenditure (EE).
Methods: Thirty participants (mean ± SD: age, 23.0 ± 2.3 yr; body mass index, 25.2 ± 3.9 kg·m) volunteered to complete the study. Participants wore five GT9X monitors (right hip, both wrists, and both ankles) while performing 10 activities ranging from rest to running. A Cosmed K4b was worn during the trial, as a criterion measure of EE (30-s averages) expressed in METs. Triaxial accelerometer data (80 Hz) were converted to milli-G using Euclidean norm minus one (ENMO; 1-s epochs). Gyroscope data (100 Hz) were expressed as a vector magnitude (GVM) in degrees per second (1-s epochs) and magnetometer data (100 Hz) were expressed as direction changes per 5 s. Minutes 4-6 of each activity were used for analysis. Three two-regression algorithms were developed for each wear location: 1) ENMO, 2) ENMO and GVM, and 3) ENMO, GVM, and direction changes. Leave-one-participant-out cross-validation was used to evaluate the root mean square error (RMSE) and mean absolute percent error (MAPE) of each algorithm.
Results: Adding gyroscope to accelerometer-only algorithms resulted in RMSE reductions between 0.0 METs (right wrist) and 0.17 METs (right ankle), and MAPE reductions between 0.1% (right wrist) and 6.0% (hip). When direction changes were added, RMSE changed by ≤0.03 METs and MAPE by ≤0.21%.
Conclusions: The combined use of gyroscope and accelerometer at the hip and ankles improved individual-level prediction of EE compared with accelerometer only. For the wrists, adding gyroscope produced negligible changes. The magnetometer did not meaningfully improve estimates for any algorithms.
Similar articles
-
Intensity Thresholds on Raw Acceleration Data: Euclidean Norm Minus One (ENMO) and Mean Amplitude Deviation (MAD) Approaches.PLoS One. 2016 Oct 5;11(10):e0164045. doi: 10.1371/journal.pone.0164045. eCollection 2016. PLoS One. 2016. PMID: 27706241 Free PMC article.
-
A random forest classifier for the prediction of energy expenditure and type of physical activity from wrist and hip accelerometers.Physiol Meas. 2014 Nov;35(11):2191-203. doi: 10.1088/0967-3334/35/11/2191. Epub 2014 Oct 23. Physiol Meas. 2014. PMID: 25340969 Free PMC article.
-
Estimating Sedentary Time from a Hip- and Wrist-Worn Accelerometer.Med Sci Sports Exerc. 2020 Jan;52(1):225-232. doi: 10.1249/MSS.0000000000002099. Med Sci Sports Exerc. 2020. PMID: 31343523
-
Comparison of linear and non-linear models for predicting energy expenditure from raw accelerometer data.Physiol Meas. 2017 Feb;38(2):343-357. doi: 10.1088/1361-6579/38/2/343. Epub 2017 Jan 20. Physiol Meas. 2017. PMID: 28107205
-
Moving Forward with Backward Compatibility: Translating Wrist Accelerometer Data.Med Sci Sports Exerc. 2016 Nov;48(11):2142-2149. doi: 10.1249/MSS.0000000000001015. Med Sci Sports Exerc. 2016. PMID: 27327029
Cited by 5 articles
-
Feasibility of the Energy Expenditure Prediction for Athletes and Non-Athletes from Ankle-Mounted Accelerometer and Heart Rate Monitor.Sci Rep. 2020 Jun 1;10(1):8816. doi: 10.1038/s41598-020-65713-7. Sci Rep. 2020. PMID: 32483254 Free PMC article.
-
Gestational weight gain and optimal wellness (GLOW): rationale and methods for a randomized controlled trial of a lifestyle intervention among pregnant women with overweight or obesity.BMC Pregnancy Childbirth. 2019 Apr 30;19(1):145. doi: 10.1186/s12884-019-2293-8. BMC Pregnancy Childbirth. 2019. PMID: 31039753 Free PMC article.
-
Estimating Physical Activity in Children Aged 8-11 Years Using Accelerometry: Contributions From Fundamental Movement Skills and Different Accelerometer Placements.Front Physiol. 2019 Mar 18;10:242. doi: 10.3389/fphys.2019.00242. eCollection 2019. Front Physiol. 2019. PMID: 30936837 Free PMC article.
-
Reliability and validity of a new accelerometer-based device for detecting physical activities and energy expenditure.PeerJ. 2018 Oct 11;6:e5775. doi: 10.7717/peerj.5775. eCollection 2018. PeerJ. 2018. PMID: 30324032 Free PMC article.
-
Estimating physical activity in youth using an ankle accelerometer.J Sports Sci. 2018 Oct;36(19):2265-2271. doi: 10.1080/02640414.2018.1449091. Epub 2018 Mar 8. J Sports Sci. 2018. PMID: 29517959 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources