Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices
- PMID: 29281579
- PMCID: PMC5894885
- DOI: 10.1056/NEJMoa1604267
Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices
Abstract
Background: Patients who have pacemakers or defibrillators are often denied the opportunity to undergo magnetic resonance imaging (MRI) because of safety concerns, unless the devices meet certain criteria specified by the Food and Drug Administration (termed "MRI-conditional" devices).
Methods: We performed a prospective, nonrandomized study to assess the safety of MRI at a magnetic field strength of 1.5 Tesla in 1509 patients who had a pacemaker (58%) or an implantable cardioverter-defibrillator (42%) that was not considered to be MRI-conditional (termed a "legacy" device). Overall, the patients underwent 2103 thoracic and nonthoracic MRI examinations that were deemed to be clinically necessary. The pacing mode was changed to asynchronous mode for pacing-dependent patients and to demand mode for other patients. Tachyarrhythmia functions were disabled. Outcome assessments included adverse events and changes in the variables that indicate lead and generator function and interaction with surrounding tissue (device parameters).
Results: No long-term clinically significant adverse events were reported. In nine MRI examinations (0.4%; 95% confidence interval, 0.2 to 0.7), the patient's device reset to a backup mode. The reset was transient in eight of the nine examinations. In one case, a pacemaker with less than 1 month left of battery life reset to ventricular inhibited pacing and could not be reprogrammed; the device was subsequently replaced. The most common notable change in device parameters (>50% change from baseline) immediately after MRI was a decrease in P-wave amplitude, which occurred in 1% of the patients. At long-term follow-up (results of which were available for 63% of the patients), the most common notable changes from baseline were decreases in P-wave amplitude (in 4% of the patients), increases in atrial capture threshold (4%), increases in right ventricular capture threshold (4%), and increases in left ventricular capture threshold (3%). The observed changes in lead parameters were not clinically significant and did not require device revision or reprogramming.
Conclusions: We evaluated the safety of MRI, performed with the use of a prespecified safety protocol, in 1509 patients who had a legacy pacemaker or a legacy implantable cardioverter-defibrillator system. No long-term clinically significant adverse events were reported. (Funded by Johns Hopkins University and the National Institutes of Health; ClinicalTrials.gov number, NCT01130896 .).
Conflict of interest statement
No other potential conflict of interest relevant to this article was reported.
Comment in
-
Imaging: Pacemakers, ICDs, and MRI.Nat Rev Cardiol. 2018 Mar;15(3):136. doi: 10.1038/nrcardio.2018.2. Epub 2018 Jan 18. Nat Rev Cardiol. 2018. PMID: 29345685 No abstract available.
-
Magnetic Resonance Imaging and Cardiac Devices.N Engl J Med. 2018 Apr 26;378(17):1652. doi: 10.1056/NEJMc1802623. N Engl J Med. 2018. PMID: 29697918 No abstract available.
-
Assessing the safety of magnetic resonance imaging in cardiac device carriers.Intern Emerg Med. 2018 Sep;13(6):933-935. doi: 10.1007/s11739-018-1876-0. Epub 2018 May 15. Intern Emerg Med. 2018. PMID: 29766404 No abstract available.
Similar articles
-
A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices.Ann Intern Med. 2011 Oct 4;155(7):415-24. doi: 10.7326/0003-4819-155-7-201110040-00004. Ann Intern Med. 2011. PMID: 21969340 Free PMC article. Clinical Trial.
-
Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator.N Engl J Med. 2017 Feb 23;376(8):755-764. doi: 10.1056/NEJMoa1603265. N Engl J Med. 2017. PMID: 28225684
-
The Safety of Cardiac and Thoracic Magnetic Resonance Imaging in Patients with Cardiac Implantable Electronic Devices.Acad Radiol. 2016 Dec;23(12):1498-1505. doi: 10.1016/j.acra.2016.08.016. Epub 2016 Oct 4. Acad Radiol. 2016. PMID: 27717762
-
Implications of pacemakers and implantable cardioverter defibrillators in urological practice.J Urol. 2011 Oct;186(4):1198-205. doi: 10.1016/j.juro.2011.02.2697. Epub 2011 Aug 16. J Urol. 2011. PMID: 21849189 Review.
-
[Magnetic resonance imaging in patients with pacemakers and implantable cardioverter-defibrillators: a systematic review].Ann Cardiol Angeiol (Paris). 2010 Aug;59(4):221-8. doi: 10.1016/j.ancard.2009.04.003. Epub 2009 May 14. Ann Cardiol Angeiol (Paris). 2010. PMID: 19962691 Review. French.
Cited by
-
Magnetic resonance-conditional cardiac implantable electronic devices: an Italian perspective on the prevalence of mixed-brand systems over time.Sci Rep. 2024 Nov 6;14(1):27006. doi: 10.1038/s41598-024-73403-x. Sci Rep. 2024. PMID: 39505923 Free PMC article.
-
Magnetic resonance imaging in patients with cardiac implantable electronic devices: the RESONANCE Spanish registry.Europace. 2024 Nov 1;26(11):euae277. doi: 10.1093/europace/euae277. Europace. 2024. PMID: 39498837 Free PMC article.
-
Assessment of Tricuspid Regurgitation by Cardiac Magnetic Resonance Imaging: Current Role and Future Applications.J Clin Med. 2024 Jul 31;13(15):4481. doi: 10.3390/jcm13154481. J Clin Med. 2024. PMID: 39124748 Free PMC article. Review.
-
Pre- and post-procedural cardiac imaging (computed tomography and magnetic resonance imaging) in electrophysiology: a clinical consensus statement of the European Heart Rhythm Association and European Association of Cardiovascular Imaging of the European Society of Cardiology.Europace. 2024 May 2;26(5):euae108. doi: 10.1093/europace/euae108. Europace. 2024. PMID: 38743765 Free PMC article. Review.
-
A Deep Learning Approach to Classify Fabry Cardiomyopathy from Hypertrophic Cardiomyopathy Using Cine Imaging on Cardiac Magnetic Resonance.Int J Biomed Imaging. 2024 Apr 26;2024:6114826. doi: 10.1155/2024/6114826. eCollection 2024. Int J Biomed Imaging. 2024. PMID: 38706878 Free PMC article.
References
-
- Gimbel JR, Johnson D, Levine PA, Wilkoff BL. Safe performance of magnetic resonance imaging on five patients with permanent cardiac pacemakers. Pacing Clin Electrophysiol. 1996;19:913–9. - PubMed
-
- Sommer T, Vahlhaus C, Lauck G, et al. MR imaging and cardiac pacemakers: invitro evaluation and in-vivo studies in 51 patients at 0.5 T. Radiology. 2000;215:869–79. - PubMed
-
- Vahlhaus C, Sommer T, Lewalter T, et al. Interference with cardiac pacemakers by magnetic resonance imaging: are there irreversible changes at 0.5 Tesla? Pacing Clin Electrophysiol. 2001;24:489–95. - PubMed
-
- Martin ET, Coman JA, Shellock FG, Pulling CC, Fair R, Jenkins K. Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla. J Am Coll Cardiol. 2004;43:1315–24. - PubMed
Publication types
MeSH terms
Associated data
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical