The use of a proximal femoral nail as a hip prosthesis: A biomechanical analysis of a newly designed implant

Proc Inst Mech Eng H. 2018 Feb;232(2):200-206. doi: 10.1177/0954411917751561. Epub 2018 Jan 3.

Abstract

Complication rates of hip fractures after proximal femoral nail are not rare. In such fractures, treatment invariably comprises the extraction of proximal femoral nail and the introduction of a different hip prosthesis. To solve this problem, we aimed to develop a new implant, the modular nail prosthesis, which combines an intramedullary nail with a hip prosthesis. The aim of this study was to determine the effect of stress load distribution on the stem using finite element analysis under laboratory conditions. For this, the shortest stem of 175 mm was chosen. In addition, six proximal femoral nail-hip prosthesis combinations were produced and tested with a biomechanical test device and passed 2300-N load bearing. According to the test results, our newly developed modular nail prosthesis can be converted to hip prosthesis securely.

Keywords: Proximal femoral nail; finite element analysis; hip prosthesis; implants.

MeSH terms

  • Biomechanical Phenomena
  • Bone Nails*
  • Femur*
  • Finite Element Analysis
  • Hip Prosthesis*
  • Mechanical Phenomena*
  • Prosthesis Design*