Multicenter evaluation of a new closed system drug-transfer device in reducing surface contamination by antineoplastic hazardous drugs

Am J Health Syst Pharm. 2018 Feb 15;75(4):199-211. doi: 10.2146/ajhp160948. Epub 2018 Jan 16.

Abstract

Purpose: Results of a study to evaluate the effectiveness of a recently introduced closed system drug-transfer device (CSTD) in reducing surface contamination during compounding and simulated administration of antineoplastic hazardous drugs (AHDs) are reported.

Methods: Wipe samples were collected from 6 predetermined surfaces in compounding and infusion areas of 13 U.S. cancer centers to establish preexisting levels of surface contamination by 2 marker AHDs (cyclophosphamide and fluorouracil). Stainless steel templates were placed over the 6 previously sampled surfaces, and the marker drugs were compounded and infused per a specific protocol using all components of the CSTD. Wipe samples were collected from the templates after completion of tasks and analyzed for both marker AHDs.

Results: Aggregated results of wipe sampling to detect preexisting contamination at the 13 study sites showed that overall, 66.7% of samples (104 of 156) had detectable levels of at least 1 marker AHD; subsequent testing after CSTD use per protocol found a sample contamination rate of 5.8% (9 of 156 samples). In the administration areas alone, the rate of preexisting contamination was 78% (61 of 78 samples); with use of the CSTD protocol, the contamination rate was 2.6%. Twenty-six participants rated the CSTD for ease of use, with 100% indicating that they were satisfied or extremely satisfied.

Conclusion: A study involving a rigorous protocol and 13 cancer centers across the United States demonstrated that the CSTD reduced surface contamination by cyclophosphamide and fluorouracil during compounding and simulated administration. Participants reported that the CSTD was easy to use.

Keywords: CSTD; antineoplastic drugs; closed system drug-transfer device; cyclophosphamide; hazardous drugs; surface contamination; wipe sampling.

Publication types

  • Evaluation Study
  • Multicenter Study

MeSH terms

  • Antineoplastic Agents / toxicity*
  • Cyclophosphamide / toxicity
  • Drug Compounding / instrumentation
  • Drug Compounding / methods
  • Drug Compounding / standards*
  • Environmental Monitoring / methods
  • Environmental Monitoring / standards*
  • Equipment Contamination / prevention & control*
  • Fluorouracil / toxicity
  • Hospitals, Community / methods
  • Hospitals, Community / standards
  • Humans
  • Occupational Exposure / prevention & control
  • Occupational Exposure / standards
  • Pharmacy Service, Hospital / methods
  • Pharmacy Service, Hospital / standards*

Substances

  • Antineoplastic Agents
  • Cyclophosphamide
  • Fluorouracil