High-sensitivity 5-, 6-, and 7-color PNH WBC assays for both Canto II and Navios platforms

Cytometry B Clin Cytom. 2018 Jul;94(4):637-651. doi: 10.1002/cyto.b.21626. Epub 2018 Mar 5.

Abstract

Background: Paroxysmal Nocturnal Hemoglobinuria (PNH) is a rare acquired hematopoietic stem cell disorder characterized by an inability to make Glyco-Phosphatidyl-Inositol (GPI)-linked cell surface structures. Fluorescent proaerolysin (FLAER-Alexa488) is increasingly used to detect GPI-deficient WBCs by flow cytometry. However, FLAER is not available in all countries and is expensive to obtain in others. An earlier study to compare FLAER-based and non-FLAER assays confirmed very good agreement between the two tubes suggesting a cost effective simultaneous evaluation of PNH neutrophils and monocytes is possible without FLAER.

Methods: We have used a single tube approach with a 7-color assay comprising FLAER-CD157-CD15-CD64-CD24-CD14-CD45. Conjugates were carefully selected and validated so that stained samples could be analyzed on either 10-color Navios or 8-color FACSCanto II platforms. The 6-color (minus CD14) and 5-color (minus CD24 and CD14) versions were also developed and compared with our predicate clinical lab 5-color assay comprising FLAER-CD157PE-CD64ECD-CD15PC5-CD45PC7.

Results/conclusions: CD15-gated PNH neutrophil clone size was quantified using either FLAER and CD157, FLAER and CD24, or CD157 and CD24. CD64-gated PNH monocyte clone size was quantified using either FLAER and CD157, FLAER and CD14, or CD157 and CD14. Analysis of >40 PNH samples showed that the FLAER-based plots derive virtually identical data to the non-FLAER plot for neutrophils (R2 = 1) and monocytes (R2 = 0.9999) and that closely similar data can be acquired using both Canto II and Navios platforms with 7-, 6-, and 5-color versions of the assay. Assessment of non-PNH samples confirmed extremely low background rate of PNH phenotypes (neutrophils and monocytes) with all three approaches. © 2018 International Clinical Cytometry Society.

Keywords: cross-platform; high sensitivity; paroxysmal nocturnal hemoglobinuria; white blood cells.

MeSH terms

  • Antigens, CD / analysis
  • Flow Cytometry / instrumentation*
  • Flow Cytometry / methods*
  • Hemoglobinuria, Paroxysmal / diagnosis*
  • Humans
  • Immunophenotyping / instrumentation*
  • Immunophenotyping / methods*
  • Sensitivity and Specificity