Preliminary experience with a transcranial magnetic resonance-guided focused ultrasound surgery system integrated with a 1.5-T MRI unit in a series of patients with essential tremor and Parkinson's disease

Neurosurg Focus. 2018 Feb;44(2):E7. doi: 10.3171/2017.11.FOCUS17614.

Abstract

OBJECTIVE Transcranial magnetic resonance-guided focused ultrasound surgery (tcMRgFUS) is one of the emerging noninvasive technologies for the treatment of neurological disorders such as essential tremor (ET), idiopathic asymmetrical tremor-dominant Parkinson's disease (PD), and neuropathic pain. In this clinical series the authors present the preliminary results achieved with the world's first tcMRgFUS system integrated with a 1.5-T MRI unit. METHODS The authors describe the results of tcMRgFUS in a sample of patients with ET and with PD who underwent the procedure during the period from January 2015 to September 2017. A monolateral ventralis intermedius nucleus (VIM) thalamic ablation was performed in both ET and PD patients. In all the tcMRgFUS treatments, a 1.5-T MRI scanner was used for both planning and monitoring the procedure. RESULTS During the study period, a total of 26 patients underwent tcMRgFUS thalamic ablation for different movement disorders. Among these patients, 18 were diagnosed with ET and 4 were affected by PD. All patients with PD were treated using tcMRgFUS thalamic ablation and all completed the procedure. Among the 18 patients with ET, 13 successfully underwent tcMRgFUS, 4 aborted the procedure during ultrasound delivery, and 1 did not undergo the tcMRgFUS procedure after stereotactic frame placement. Two patients with ET were not included in the results because of the short follow-up duration at the time of this study. A monolateral VIM thalamic ablation in both ET and PD patients was performed. All the enrolled patients were evaluated before the treatment and 2 days after, with a clinical control of the treatment effectiveness using the graphic items of the Fahn-Tolosa-Marin tremor rating scale. A global reevaluation was performed 3 months (17/22 patients) and 6 months (11/22 patients) after the treatment; the reevaluation consisted of clinical questionnaires, neurological tests, and video recordings of the tests. All the ET and PD treated patients who completed the procedure showed an immediate amelioration of tremor severity, with no intra- or posttreatment severe permanent side effects. CONCLUSIONS Although this study reports on a small number of patients with a short follow-up duration, the tcMRgFUS procedure using a 1.5-T MRI unit resulted in a safe and effective treatment option for motor symptoms in patients with ET and PD. To the best of the authors' knowledge, this is the first clinical series in which thalamotomy was performed using tcMRgFUS integrated with a 1.5-T magnet.

Keywords: BBB = blood-brain barrier; ET = essential tremor; FRFSE = fast recalled FSE; FSE = fast spin echo; FSPGR = fast spoiled gradient echo; FTM = Fahn-Tolosa-Marin; HI-FU = high-intensity focused ultrasound; MRI; MRgFUS; MS = multiple sclerosis; PD = Parkinson’s disease; Parkinson’s disease; QUEST = Quality of Life in Essential Tremor; SWAN = susceptibility-weighted angiography; UPDRS = Unified Parkinson’s Disease Rating Scale; VIM = ventralis intermedius nucleus; brain; essential tremor; high-intensity focused ultrasound ablation; interventional; magnetic resonance–guided focused ultrasound surgery; stereotactic technique; tcMRgFUS = transcranial magnetic resonance–guided focused ultrasound surgery.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Essential Tremor / diagnostic imaging*
  • Essential Tremor / surgery
  • Female
  • Follow-Up Studies
  • Humans
  • Intraoperative Neurophysiological Monitoring / methods*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • Parkinson Disease / diagnostic imaging*
  • Parkinson Disease / surgery
  • Thalamus / diagnostic imaging*
  • Thalamus / surgery
  • Treatment Outcome
  • Ultrasonography, Interventional / methods*