Three-year Outcomes of a Randomized Controlled Trial Comparing a 4.5-mm-Wide to a 3.75-mm-Wide Titanium Implant for Bone Conduction Hearing

Otol Neurotol. 2018 Jun;39(5):609-615. doi: 10.1097/MAO.0000000000001761.

Abstract

Objective: To compare 3-year implant stability, survival, and tolerability of a 4.5-mm-wide (test) and a 3.75-mm-wide (control) percutaneous titanium implant for bone-conduction hearing, loaded with the sound processor after 3 weeks.

Methods: Sixty implants were allocated in a 2:1 ratio (test-control) in 57 adult patients included in this prospective randomized controlled clinical trial. Follow-up visits were performed at 7, 14, 21, and 28 days; 6 and 12 weeks; 6 months; and at 1, 2, and 3 years after implantation. During these visits, the implant stability quotient (ISQ) was measured by means of resonance frequency analysis (RFA). The peri-abutment soft tissue status was assessed according to the Holgers classification. Skin height around the abutment was evaluated.

Results: The mean area-under-the-curve (AUC) of ISQ-low was statistically significantly higher for the test implant (65.7 versus 61.4, p = 0.0002). Both implants showed high survival rates (97.4% versus 95.0%, p = 0.6374). Adverse soft tissue reactions were observed sporadically, with no significant inter-group differences. Skin thickening was seen in the majority of the patients, but no correlation with adverse soft tissue reactions or implant type was observed.

Conclusion: The 4.5-mm-wide implant provides significantly higher ISQ values during the first 3 years after surgery compared with the previous generation 3.75-mm-wide implant. Both implants showed high survival rates and good tolerability. These long-term results indicate that the wider implant, loaded with a sound processor at 3 weeks, is a safe and well-performing option for hearing rehabilitation in specific types of hearing loss.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Bone Conduction*
  • Female
  • Hearing Aids* / adverse effects
  • Hearing Loss / surgery*
  • Humans
  • Male
  • Middle Aged
  • Osseointegration
  • Prospective Studies
  • Prostheses and Implants*
  • Titanium
  • Treatment Outcome

Substances

  • Titanium