Cardiovascular magnetic resonance assessment of 1st generation CoreValve and 2nd generation Lotus valves

J Interv Cardiol. 2018 Jun;31(3):391-399. doi: 10.1111/joic.12512. Epub 2018 Apr 25.

Abstract

Objectives: We sought to compare using serial CMR, the quantity of AR and associated valve hemodynamics, following the first-generation CoreValve (Medtronic, Minneapolis, MN) and the second-generation Lotus valve (Boston Scientific, Natick, MA).

Background: Aortic regurgitation (AR) following Transcatheter Aortic Valve Replacement (TAVR) confers a worse prognosis and can be accurately quantified using cardiovascular magnetic resonance (CMR). Second generation valves have been specifically designed to reduce paravalvular AR and improve clinical outcomes.

Methods: Fifty-one patients (79.0 ± 7.7 years, 57% male) were recruited and imaged at three time points: immediately pre- and post-TAVR, and at 6 months.

Results: CMR-derived AR fraction immediately post-TAVR was greater in the CoreValve compared to Lotus group (11.7 ± 8.4 vs. 4.3 ± 3.4%, P = 0.001), as was the frequency of ≥moderate AR (9/24 (37.5%) versus 0/27, P < 0.001). However, at 6 months AR fraction had improved significantly in the CoreValve group such that the two valve designs were comparable (6.4 ± 5.0 vs 5.6 ± 5.3%, P = 0.623), with no patient in either group having ≥moderate AR. The residual peak pressure gradient immediately following TAVR was significantly lower with CoreValve compared to Lotus (14.1 ± 5.6 vs 25.4 ± 11.6 mmHg, P = 0.001), but again by 6 months the two valve designs were comparable (16.5 ± 9.4 vs 19.7 ± 10.5 mmHg, P = 0.332). There was no difference in the degree of LV reverse remodeling between the two valves at 6 months.

Conclusion: Immediately post-TAVR, there was significantly less AR but a higher residual peak pressure gradient with the Lotus valve compared to CoreValve. However, at 6 months both devices had comparable valve hemodynamics and LV reverse remodeling.

Keywords: Boston lotus valve; aortic regurgitation; cardiovascular magnetic resonance; medtronic CoreValve; reverse remodeling; transcatheter aortic valve replacement.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / surgery
  • Aortic Valve Insufficiency / diagnostic imaging*
  • Aortic Valve Insufficiency / etiology
  • Aortic Valve Insufficiency / surgery*
  • Aortic Valve Stenosis / complications
  • Aortic Valve Stenosis / diagnostic imaging*
  • Aortic Valve Stenosis / surgery*
  • Female
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Magnetic Resonance Imaging*
  • Male
  • Prosthesis Design
  • Transcatheter Aortic Valve Replacement
  • Treatment Outcome