A novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and positioning systems
- PMID: 29730884
- DOI: 10.1002/mp.12950
A novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and positioning systems
Abstract
Purpose: Various localization and positioning systems utilizing radiographic or nonradiographic methods have been developed to improve the accuracy of radiation treatment. Each quality assurance (QA) procedure requires its own phantom and is independent from each other, so the deviation between each system is unavailable. The purpose of this work is to develop and evaluate a single-integrated QA phantom for different localization and positioning systems.
Methods: The integrated phantom was designed in three-dimensional (3D) CAD software and 3D printed. The phantom was designed with laser alignment marks, a raised letter "S" on the anterior surface for optical surface monitoring system registration, a core for radiofrequency (RF) tracking system alignment, eight internal fiducials for image alignment, and an isocentric bearing for Winston-Lutz test. Tilt legs and rotational stage were designed for rotational verification of optical surface mapping system and RF tracking system, respectively. The phantom was scanned using a CT scanner and a QA plan was created. This prototype phantom was evaluated against established QA techniques.
Results: The QA result between the proposed procedure and established QA technique are 1.12 ± 0.31 and 1.14 ± 0.31 mm, respectively, for RF tracking system and 0.18 ± 0.06 and 0.18 ± 0.05 mm for Winston-Lutz test. There is no significant difference for the QA results between the established QA and proposed procedure (P > 0.05, t test). The accuracy of rotational verification for surface mapping system and RF tracking system are less than 0.5 and 1° compared the predefined value. The isocenter deviation of each location system is around l mm.
Conclusion: We have designed and evaluated a novel-integrated phantom for radiographic and nonradiographic localization and positioning systems for radiotherapy. With this phantom, we will reduce the variation in measurements and simplify the QA procedures.
Keywords: localization positioning system; nonradiographic and radiographic QA; quality assurance; radiotherapy.
© 2018 American Association of Physicists in Medicine.
Similar articles
-
Quality assurance for a six degrees-of-freedom table using a 3D printed phantom.J Appl Clin Med Phys. 2018 Jan;19(1):115-124. doi: 10.1002/acm2.12227. Epub 2017 Nov 21. J Appl Clin Med Phys. 2018. PMID: 29159920 Free PMC article.
-
An integrated quality assurance phantom for frameless single-isocenter multitarget stereotactic radiosurgery.Phys Med Biol. 2020 Jun 10;65(11):115006. doi: 10.1088/1361-6560/ab8534. Phys Med Biol. 2020. PMID: 32235050
-
A novel phantom and procedure providing submillimeter accuracy in daily QA tests of accelerators used for stereotactic radiosurgery*.J Appl Clin Med Phys. 2016 Jul 8;17(4):246-253. doi: 10.1120/jacmp.v17i4.6295. J Appl Clin Med Phys. 2016. PMID: 27455506 Free PMC article.
-
Quality assurance of ultrasound systems: current status and review of literature.J Ultrasound. 2018 Sep;21(3):173-182. doi: 10.1007/s40477-018-0304-7. Epub 2018 Jun 14. J Ultrasound. 2018. PMID: 29949119 Free PMC article. Review.
-
Quality assurance of human functional magnetic resonance imaging: a literature review.Quant Imaging Med Surg. 2019 Jun;9(6):1147-1162. doi: 10.21037/qims.2019.04.18. Quant Imaging Med Surg. 2019. PMID: 31367569 Free PMC article. Review.
Cited by
-
Development and Longitudinal Analysis of Plan-Based Streamlined Quality Assurance on Multiple Positioning Guidance Systems With Single Phantom Setup.Front Oncol. 2021 Jun 16;11:683733. doi: 10.3389/fonc.2021.683733. eCollection 2021. Front Oncol. 2021. PMID: 34222005 Free PMC article.
-
Recent advanced in Surface Guided Radiation Therapy.Radiat Oncol. 2020 Jul 31;15(1):187. doi: 10.1186/s13014-020-01629-w. Radiat Oncol. 2020. PMID: 32736570 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous