Simultaneous T1 and T2 mapping of the carotid plaque (SIMPLE) with T2 and inversion recovery prepared 3D radial imaging

Magn Reson Med. 2018 Dec;80(6):2598-2608. doi: 10.1002/mrm.27361. Epub 2018 May 25.

Abstract

Purpose: To propose a technique that can produce different T1 and T2 contrasts in a single scan for simultaneous T1 and T2 mapping of the carotid plaque (SIMPLE).

Methods: An interleaved 3D golden angle radial trajectory was used in conjunction with T2 preparation with variable duration (TEprep ) and inversion recovery pulses. Sliding window reconstruction was adopted to reconstruct images at different inversion delay time and TEprep for joint T1 and T2 fitting. In the fitting procedure, a rapid B1 correction method was presented. The accuracy of SIMPLE was investigated in phantom experiments. In vivo scans were performed on 5 healthy volunteers with 2 scans each, and on 5 patients with carotid atherosclerosis.

Results: The phantom T1 and T2 estimations of SIMPLE agreed well with the standard methods with the percentage difference smaller than 7.1%. In vivo T1 and T2 for normal carotid vessel wall were 1213 ± 48.3 ms and 51.1 ± 1.7 ms, with good interscan repeatability. Alternations of T1 and T2 in plaque regions were in agreement with the conventional multicontrast imaging findings.

Conclusion: The proposed SIMPLE allows simultaneous T1 and T2 mapping of the carotid artery in less than 10 minutes, serving as a quantitative tool with good accuracy and reproducibility for plaque characterization.

Keywords: 3D radial; T1 mapping; T2 mapping; atherosclerosis; carotid artery.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Carotid Arteries / diagnostic imaging
  • Carotid Artery Diseases / diagnostic imaging*
  • Carotid Stenosis / diagnostic imaging*
  • Contrast Media
  • Female
  • Humans
  • Image Interpretation, Computer-Assisted / methods
  • Image Processing, Computer-Assisted
  • Imaging, Three-Dimensional / methods*
  • Male
  • Middle Aged
  • Phantoms, Imaging
  • Reproducibility of Results
  • Young Adult

Substances

  • Contrast Media