Comparison of a Simplified Skin Pointer Device Compared With a Skeletal Marker for Knee Rotation Laxity: A Cadaveric Study Using a Rotation-Meter

Affiliations

Comparison of a Simplified Skin Pointer Device Compared With a Skeletal Marker for Knee Rotation Laxity: A Cadaveric Study Using a Rotation-Meter

Ken Lee Puah et al. World J Orthop.

Abstract

Aim: To compare the measurements of knee rotation laxity by non-invasive skin pointer with a knee rotation jig in cadaveric knees against a skeletally mounted marker.

Methods: Six pairs of cadaveric legs were mounted on a knee rotation jig. One Kirscher wire was driven into the tibial tubercle as a bone marker and a skin pointer was attached. Rotational forces of 3, 6 and 9 nm applied at 0°, 30°, 45°, 60° and 90° of knee flexion were analysed using the Pearson correlation coefficient and paired t-test.

Results: Total rotation recorded with the skin pointer significantly correlated with the bone marker at 3 nm at 0° (skin pointer 23.9 ± 26.0° vs bone marker 16.3 ± 17.3°, r = 0.92; P = 0.0), 30° (41.7 ± 15.5° vs 33.1 ± 14.7°, r = 0.63; P = 0.037), 45° (49.0 ± 17.0° vs 40.3 ± 11.2°, r = 0.81; P = 0.002), 60° (45.7 ± 17.5° vs 34.7 ± 9.5°, r = 0.86; P = 0.001) and 90° (29.2 ± 10.9° vs 21.2 ± 6.8°, r = 0.69; P = 0.019) of knee flexion and 6 nm at 0° (51.1 ± 37.7° vs 38.6 ± 30.1°, r = 0.90; P = 0.0), 30° (64.6 ± 21.6° vs 54.3 ± 15.1°, r = 0.73; P = 0.011), 45° (67.7 ± 20.6° vs 55.5 ± 9.5°, r = 0.65; P = 0.029), 60° (62.9 ± 22.4° vs 45.8 ± 13.1°, r = 0.65; P = 0.031) and 90° (43.6 ± 17.6° vs 31.0 ± 6.3°, r = 0.62; P = 0.043) of knee flexion and at 9 nm at 0° (69.7 ± 40.0° vs 55.6 ± 30.6°, r = 0.86; P = 0.001) and 60° (74.5 ± 27.6° vs 57.1 ± 11.5°, r = 0.77; P = 0.006). No statistically significant correlation with 9 nm at 30° (79.2 ± 25.1° vs 66.9 ± 15.4°, r = 0.59; P = 0.055), 45° (80.7 ± 24.7° vs 65.5 ± 11.2°, r = 0.51; P = 0.11) and 90° (54.7 ± 21.1° vs 39.4 ± 8.2°, r = 0.55; P = 0.079). We recognize that 9 nm of torque may be not tolerated in vivo due to pain. Knee rotation was at its maximum at 45° of knee flexion and increased with increasing torque.

Conclusion: The skin pointer and knee rotation jig can be a reliable and simple means of quantifying knee rotational laxity with future clinical application.

Keywords: Biomechanics; Cruciate; Knee; Laxity; Measurement; Rotatometer; Rottometer.

Conflict of interest statement

Conflict-of-interest statement: No potential conflicts of interest relevant to this article were reported.

Figures

Figure 1
Figure 1
Knee rotation jig prior to mounting of specimen.
Figure 2
Figure 2
Knee rotation jig with specimen mounted with nail though tibial tuberosity and skin pointer in place.
Figure 3
Figure 3
Knee rotation measured at 0°, 30°, 45°, 60° and 90° of knee flexion with 3 nm of torque.
Figure 4
Figure 4
Knee rotation measured at 0°, 30°, 45°, 60° and 90° of knee flexion with 6 nm of torque.
Figure 5
Figure 5
Knee rotation measured at 0°, 30°, 45°, 60° and 90° of knee flexion with 9 nm of torque.
Figure 6
Figure 6
Knee rotation at 45° of flexion with 3, 6, and 9 nm of torque.

Similar articles

See all similar articles

References

    1. Ahldén M, Samuelsson K, Sernert N, Forssblad M, Karlsson J, Kartus J. The Swedish National Anterior Cruciate Ligament Register: a report on baseline variables and outcomes of surgery for almost 18,000 patients. Am J Sports Med. 2012;40:2230–2235. - PubMed
    1. Araki D, Kuroda R, Kubo S, Fujita N, Tei K, Nishimoto K, Hoshino Y, Matsushita T, Matsumoto T, Nagamune K, et al. A prospective randomised study of anatomical single-bundle versus double-bundle anterior cruciate ligament reconstruction: quantitative evaluation using an electromagnetic measurement system. Int Orthop. 2011;35:439–446. - PMC - PubMed
    1. Kärrholm J. Roentgen stereophotogrammetry. Review of orthopedic applications. Acta Orthop Scand. 1989;60:491–503. - PubMed
    1. Selvik G. Roentgen stereophotogrammetry. A method for the study of the kinematics of the skeletal system. Acta Orthop Scand Suppl. 1989;232:1–51. - PubMed
    1. Musahl V, Hoshino Y, Ahlden M, Araujo P, Irrgang JJ, Zaffagnini S, Karlsson J, Fu FH. The pivot shift: a global user guide. Knee Surg Sports Traumatol Arthrosc. 2012;20:724–731. - PubMed
Feedback