Are three-dimensional patient-specific cutting guides for open wedge high tibial osteotomy accurate? An in vitro study

J Orthop Surg Res. 2018 Jul 9;13(1):171. doi: 10.1186/s13018-018-0872-4.

Abstract

Background: The aim of this in vitro study was to assess the accuracy of three-dimensional patient-specific cutting guides for open wedge high tibial osteotomy (OWHTO) to provide the planned correction in both frontal and sagittal planes.

Methods: Ten cadaveric tibias underwent OWHTO performed using a patient-specific cutting guide based on 3D preoperative planning. An initial CT scan of the tibias was performed, and after segmentation, 3D geometrical models of the pre-OWHTO tibias were obtained. Reference planes were defined, and OWHTO virtually planned to then design patient-specific cutting guides. OWHTO were performed using the patient-specific cutting guides. The patient-specific cutting guide controls the cut and the correction of the OWHTO in both planes. 3D models of post-OWHTO tibias were created after a postoperative CT scan. Geometrical post-OWHTO 3D models were superimposed on pre-OWHTO 3D models. Mechanical medial proximal tibial angle (mMPTA) in the frontal plane and posterior tibial slope (PTS) in the sagittal plane were compared between planned-OWHTO and post-OWHTO 3D reconstructions relative to the pre-OWHTO reference planes and axis. Pearson's and Lin's correlation tests were performed to assess precision and accuracy of patient-specific cutting guides.

Results: The mean difference between post-OWHTO and planned-OWHTO was 0.2° (max 0.5°, SD 0.3°) in the frontal plane and - 0.1° (max 0.8°, SD 0.5°) in the sagittal plane. Statistically significant correlations were found between the planned-OWHTO and post-OWHTO configurations for the mMPTA (p < 0.0001) and PTS (p < 0.0001) measurements, and the bias correction factor was 0.99 in both planes.

Conclusions: 3D patient-specific cutting guides for OWHTO-based 3D virtual planning is a reliable and accurate method of achieving multiplanar correction in both frontal and sagittal planes.

Keywords: Accuracy; Knee surgery; Medial gonarthrosis; Open wedge high tibial osteotomy; Osteoarthrosis; Osteotomy; Patient-specific; Tibial slope correction.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Cadaver
  • Female
  • Humans
  • Imaging, Three-Dimensional
  • Male
  • Osteoarthritis, Knee / diagnostic imaging
  • Osteoarthritis, Knee / surgery*
  • Osteotomy / instrumentation*
  • Osteotomy / methods
  • Surgery, Computer-Assisted
  • Tibia / surgery*
  • Tomography, X-Ray Computed