Increased separability of K-edge nanoparticles by photon-counting detectors for spectral micro-CT

J Xray Sci Technol. 2018;26(5):707-726. doi: 10.3233/XST-18382.

Abstract

Background: X-ray CT/micro-CT methods with photon-counting detectors (PCDs) and high Z materials are a hot research topic. One method using PCDs allows for spectral imaging in 5 energy windows while conventional X-ray detectors only collect energy-integrating data.

Objective: To demonstrate the enhanced separation of contrast materials by using PCDs, multivariate analysis, and linear discriminant methods.

Methods: Phantoms containing iodine and aqueous nanomaterials were scanned on a MARS spectral micro-CT. Image volumes were segmented into separate material-specific populations. Contrast comparisons were made by calculating T2 test statistics in the univariate, pseudo-conventional and multivariate, spectral CT data sets. Separability after Fisher discriminant analysis (FDA) was also assessed.

Results: The T2 values calculated for material comparisons increased as a result of the spectral expansion. The majority of the tested contrast agents showed increased T2 values by a factor of ∼2 -3. The total significant T2 statistics in the pure and mixed lanthanide image sets increased in the spectral data set.

Conclusion: This work consolidates the groundwork for photon-counting-based material decomposition with micro-CT, facilitating future development of novel nanomaterials and their preclinical applications.

Keywords: K-edge imaging; Material-discriminative performance; dimensionality reduction; multivariate statistics; spectral CT.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Contrast Media
  • Equipment Design
  • Iodine
  • Nanoparticles / chemistry*
  • Phantoms, Imaging
  • Photons
  • X-Ray Microtomography / instrumentation*
  • X-Ray Microtomography / methods*

Substances

  • Contrast Media
  • Iodine