Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2018 Oct;137(2):79-86.
doi: 10.1007/s10633-018-9652-z. Epub 2018 Jul 25.

Effect of varying skin surface electrode position on electroretinogram responses recorded using a handheld stimulating and recording system

Affiliations

Effect of varying skin surface electrode position on electroretinogram responses recorded using a handheld stimulating and recording system

Angharad E Hobby et al. Doc Ophthalmol. 2018 Oct.

Abstract

Purpose: A handheld device (the RETeval system, LKC Technologies) aims to increase the ease of electroretinogram (ERG) recording by using specially designed skin electrodes, rather than corneal electrodes. We explored effects of electrode position on response parameters recorded using this device.

Methods: Healthy adult twins were recruited from the TwinsUK cohort and underwent recording of light-adapted flicker ERGs (corresponding to international standard stimuli). In Group 1, skin electrodes were placed in a "comfortable" position, which was up to 20 mm below the lid margin. For subsequent participants (Group 2), the electrode was positioned 2 mm from the lid margin as recommended by the manufacturer. Amplitudes and peak times (averaged from both eyes) were compared between groups after age-matching and inclusion of only one twin per pair. Light-adapted flicker and flash ERGs were recorded for an additional 10 healthy subjects in two consecutive recording sessions: in the test eye, electrode position was varied from 2 to 10-20 mm below the lid margin between sessions; in the fellow (control) eye, the electrode was 2 mm below the lid margin throughout. Amplitudes and peak times (test eye normalised to control eye) were compared for the two sessions.

Results: Including one twin per pair, and age-matching yielded 28 individuals per group. Flicker ERG amplitudes were significantly lower for Group 1 than Group 2 participants (p = 0.0024). However, mean peak times did not differ between groups (p = 0.54). For the subjects in whom electrode position was changed between recording sessions, flash and flicker amplitudes were significantly lower when positioned further from the lid margin (p < 0.005), but peak times were similar (p > 0.5).

Conclusions: Moving the skin electrodes further from the lid margin significantly reduces response amplitudes, highlighting the importance of consistent electrode positioning. However, this does not significantly affect peak times. Thus, it may be feasible to adopt a more comfortable position in participants who cannot tolerate the recommended position if analysis is restricted to peak time parameters.

Keywords: Electrode; Electroretinogram; Retina; Retinal function.

PubMed Disclaimer

Conflict of interest statement

Conflicts of interest

The authors declare that they have no conflict of interest.

Informed consent

All participants gave informed consent.

Statement of human rights

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.

Statement on the welfare of animals

No animals were used in this research.

Figures

Fig. 1
Fig. 1
Two different electrode positions in a study participant. The skin sensor electrodes are placed 2 mm below the lid margin (manufacturer’s recommended position) in the upper photograph, and in an inferior position 10–20 mm below lid margin (lower photograph). Permission granted for use of photograph
Fig. 2
Fig. 2
Boxplots comparing the distribution of ERG parameters between participants in Groups 1 and 2. In Group 1, the electrode was in a comfortable position (up to 20 mm from the lid margin); in Group 2, the electrode was placed in the recommended position of 2 mm below the lid margin. Median, upper and lower quartiles (limits of box) are shown. Squares plot means; whiskers span the 5th–95th centiles; and crosses plot minimum and maximum data points. Amplitudes are plotted in a and peak times in b
Fig. 3
Fig. 3
Sample traces from one participant. a, b Averaged light-adapted ERGs to a flash obtained from right and left eyes, respectively. In the right eye, the skin electrode was placed 2 mm from the lid margin. In the left eye, the electrode was 15 mm below the lid margin. c, d Light-adapted responses to the flicker stimulus obtained with electrodes placed as in a, b, respectively. Green and orange traces represent averaged recordings obtained from consecutive runs of stimulus presentations
Fig. 4
Fig. 4
ERG parameters for different electrode positions in participants in the third group (n = 10; data normalised to control eye). a, b, c Boxplots showing a-wave, b-wave and flicker ERG amplitudes: differences were significant between the two positions (p = 0.004, 0.002 and 0.002, respectively, Wilcoxon signed rank test). d, e, f, Boxplots showing a-wave, b-wave and flicker ERG peak times: differences were not significant (p = 0.85, 0.57 and 0.56, respectively). Boxes show median and upper and lower quartiles; squares plot mean values; whiskers extend to maximum and minimum values. Note that the y-axis scales differ in the lower panels and have been expanded considerably in F, where the actual range between maximum and minimum values is less than 0.1

Similar articles

Cited by

References

    1. Maa AY, Feuer WJ, Davis CQ, Pillow EK, Brown TD, Caywood RM, Chasan JE, Fransen SR. A novel device for accurate and efficient testing for vision-threatening diabetic retinopathy. J Diabetes Complications. 2016;30(3):524–532. doi: 10.1016/j.jdiacomp.2015.12.005. - DOI - PMC - PubMed
    1. Fukuo M, Kondo M, Hirose A, Fukushima H, Ikesugi K, Sugimoto M, Kato K, Uchigata Y, Kitano S. Screening for diabetic retinopathy using new mydriasis-free, full-field flicker ERG recording device. Sci Rep. 2016;8(6):36591. doi: 10.1038/srep36591. - DOI - PMC - PubMed
    1. Nakamura N, Fujinami K, Mizuno Y, Noda T, Tsunoda K. Evaluation of cone function by a handheld non-mydriatic flicker electroretinogram device. Clin Ophthalmol. 2016;30(10):1175–1185. doi: 10.2147/OPTH.S104721. - DOI - PMC - PubMed
    1. Motokawa K, Mita T. Uber eine einfachere Untersuchungsmethode und Eigenschaften der Aktionsstrome der Netzhaut des Menschen. Tohokku J Exp Med. 1942;42:114–133. doi: 10.1620/tjem.42.114. - DOI
    1. Noonan BD, Wilkus RJ, Chatrian GE, Lettich E. The influence of direction of gaze on the human electroretinogram recorded from periorbital electrodes: a study utilizing a summating technique. Electroencephalogr Clin Neurophysiol. 1973;35(5):495–502. doi: 10.1016/0013-4694(73)90025-4. - DOI - PubMed

Publication types

LinkOut - more resources