Neuromonitoring During Robotic Cochlear Implantation: Initial Clinical Experience

Ann Biomed Eng. 2018 Oct;46(10):1568-1581. doi: 10.1007/s10439-018-2094-7. Epub 2018 Jul 26.

Abstract

During robotic cochlear implantation a drill trajectory often passes at submillimeter distances from the facial nerve due to close lying critical anatomy of the temporal bone. Additional intraoperative safety mechanisms are thus required to ensure preservation of this vital structure in case of unexpected navigation system error. Electromyography based nerve monitoring is widely used to aid surgeons in localizing vital nerve structures at risk of injury during surgery. However, state of the art neuromonitoring systems, are unable to discriminate facial nerve proximity within submillimeter ranges. Previous work demonstrated the feasibility of utilizing combinations of monopolar and bipolar stimulation threshold measurements to discretize facial nerve proximity with greater sensitivity and specificity, enabling discrimination between safe (> 0.4 mm) and unsafe (< 0.1 mm) trajectories during robotic cochlear implantation (in vivo animal model). Herein, initial clinical validation of the determined stimulation protocol and nerve proximity analysis integrated into an image guided system for safety measurement is presented. Stimulation thresholds and corresponding nerve proximity values previously determined from an animal model have been validated in a first-in-man clinical trial of robotic cochlear implantation. Measurements performed automatically at preoperatively defined distances from the facial nerve were used to determine safety of the drill trajectory intraoperatively. The presented system and automated analysis correctly determined sufficient safety distance margins (> 0.4 mm) to the facial nerve in all cases.

Keywords: Accuracy; Bipolar; Electromyography; Facial nerve; Image-guided; Multipolar; Safety; Stimulation; Submillimeter.

Publication types

  • Clinical Trial

MeSH terms

  • Adult
  • Cochlear Implantation* / instrumentation
  • Cochlear Implantation* / methods
  • Cochlear Implants*
  • Electromyography* / instrumentation
  • Electromyography* / mortality
  • Female
  • Humans
  • Male
  • Monitoring, Physiologic / instrumentation
  • Monitoring, Physiologic / methods
  • Robotic Surgical Procedures* / instrumentation
  • Robotic Surgical Procedures* / methods