Ultrathin-strut, biodegradable-polymer, sirolimus-eluting stents versus thin-strut, durable-polymer, everolimus-eluting stents for percutaneous coronary revascularisation: 5-year outcomes of the BIOSCIENCE randomised trial
- PMID: 30170848
- DOI: 10.1016/S0140-6736(18)31715-X
Ultrathin-strut, biodegradable-polymer, sirolimus-eluting stents versus thin-strut, durable-polymer, everolimus-eluting stents for percutaneous coronary revascularisation: 5-year outcomes of the BIOSCIENCE randomised trial
Abstract
Background: Drug-eluting stents combining an ultrathin cobalt-chromium stent platform with a biodegradable polymer eluting sirolimus have been shown to be non-inferior or superior to thin-strut, durable-polymer, everolimus-eluting stents in terms of 1 year safety and efficacy outcomes.
Methods: In the randomised, single-blind, multicentre, non-inferiority BIOSCIENCE trial, we compared biodegradable-polymer sirolimus-eluting stents with durable-polymer everolimus-eluting stents in patients with chronic stable coronary artery disease or acute coronary syndromes. Here, we assess the final 5-year clinical outcomes of BIOSCIENCE with regards to the primary clinical outcome of target lesion failure, which was a composite of cardiac death, target vessel myocardial infarction, and clinically indicated target lesion revascularisation. The primary analysis was done by intention to treat. The BIOSCIENCE trial is registered with ClinicalTrials.gov, number NCT01443104.
Findings: 2008 (95%) of 2119 patients recruited between March 1, 2012, and May 31, 2013, completed 5 years of follow-up. Target lesion failure occurred in 198 patients (cumulative incidence 20·2%) treated with biodegradable-polymer sirolimus-eluting stents and in 189 patients (18·8%) treated with durable-polymer everolimus-eluting stents (rate ratio [RR] 1·07, 95% CI 0·88-1·31; p=0·487). All-cause mortality was significantly higher in patients treated with biodegradable-polymer sirolimus-eluting stents than in those treated with durable-polymer everolimus-eluting stents (14·1% vs 10·3%; RR 1·36, 95% CI 1·06-1·75; p=0·017), driven by a difference in non-cardiovascular deaths. We observed no difference between groups in cumulative incidence of definite stent thrombosis at 5 years (1·6% in both groups; 1·02, 0·51-2·05; p=0·950).
Interpretation: 5-year risk of target lesion failure among all-comer patients undergoing percutaneous coronary intervention is similar after implantation of ultrathin-strut, biodegradable-polymer, sirolimus-eluting stents or thin-strut, durable-polymer, everolimus-eluting stents. Higher incidences of all-cause and non-cardiovascular mortality in patients treated with biodegradable-polymer stents eluting sirolimus than in those treated with durable-polymer stents eluting everolimus warrant careful observation in ongoing clinical trials.
Funding: Clinical Trials Unit of the University of Bern and Biotronik.
Copyright © 2018 Elsevier Ltd. All rights reserved.
Comment in
-
Late clinical outcome of stent trials: a matter of life or death?Lancet. 2018 Sep 1;392(10149):713-714. doi: 10.1016/S0140-6736(18)31860-9. Epub 2018 Aug 28. Lancet. 2018. PMID: 30170847 No abstract available.
-
Biodegradable-polymer stents versus durable-polymer stents.Lancet. 2019 May 11;393(10184):1932-1933. doi: 10.1016/S0140-6736(19)30023-6. Lancet. 2019. PMID: 31084958 No abstract available.
Similar articles
-
Ultrathin strut biodegradable polymer sirolimus-eluting stent versus durable polymer everolimus-eluting stent for percutaneous coronary revascularisation (BIOSCIENCE): a randomised, single-blind, non-inferiority trial.Lancet. 2014 Dec 13;384(9960):2111-22. doi: 10.1016/S0140-6736(14)61038-2. Epub 2014 Sep 1. Lancet. 2014. PMID: 25189359 Clinical Trial.
-
Biodegradable polymer sirolimus-eluting stents versus durable polymer everolimus-eluting stents in patients with ST-segment elevation myocardial infarction (BIOSTEMI): a single-blind, prospective, randomised superiority trial.Lancet. 2019 Oct 5;394(10205):1243-1253. doi: 10.1016/S0140-6736(19)31877-X. Epub 2019 Sep 2. Lancet. 2019. PMID: 31488372 Clinical Trial.
-
Very thin strut biodegradable polymer everolimus-eluting and sirolimus-eluting stents versus durable polymer zotarolimus-eluting stents in allcomers with coronary artery disease (BIO-RESORT): a three-arm, randomised, non-inferiority trial.Lancet. 2016 Nov 26;388(10060):2607-2617. doi: 10.1016/S0140-6736(16)31920-1. Epub 2016 Oct 30. Lancet. 2016. PMID: 27806902 Clinical Trial.
-
Safety and efficacy of ultrathin strut biodegradable polymer sirolimus-eluting stent versus durable polymer drug-eluting stents: a meta-analysis of randomized trials.BMC Cardiovasc Disord. 2018 Aug 15;18(1):170. doi: 10.1186/s12872-018-0902-5. BMC Cardiovasc Disord. 2018. PMID: 30111289 Free PMC article. Review.
-
Biodegradable polymer sirolimus-eluting stents vs durable polymer everolimus-eluting stents in patients undergoing percutaneous coronary intervention: A meta-analysis of individual patient data from 5 randomized trials.Am Heart J. 2021 May;235:140-148. doi: 10.1016/j.ahj.2021.02.009. Epub 2021 Feb 17. Am Heart J. 2021. PMID: 33609498 Review.
Cited by
-
Cost-Effectiveness of Three Different New-Generation Drug-Eluting Stents in the Randomised BIO-RESORT Trial at 3 Years.Pharmacoecon Open. 2024 Oct 29. doi: 10.1007/s41669-024-00539-x. Online ahead of print. Pharmacoecon Open. 2024. PMID: 39472363
-
Safety and Efficacy of the Supreme Biodegradable Polymer Sirolimus-Eluting Stent in Patients With Diabetes Mellitus.J Soc Cardiovasc Angiogr Interv. 2022 Apr 11;1(2):100033. doi: 10.1016/j.jscai.2022.100033. eCollection 2022 Mar-Apr. J Soc Cardiovasc Angiogr Interv. 2022. PMID: 39132558 Free PMC article.
-
Comparative effectiveness of ultrathin vs. standard strut drug-eluting stents: insights from a large-scale meta-analysis with extended follow-up.Eur J Med Res. 2024 Jul 27;29(1):388. doi: 10.1186/s40001-024-01949-7. Eur J Med Res. 2024. PMID: 39068447 Free PMC article.
-
Drug-Coated Balloon Treatment for De Novo Coronary Lesions: Current Status and Future Perspectives.Korean Circ J. 2024 Sep;54(9):519-533. doi: 10.4070/kcj.2024.0148. Epub 2024 Jun 3. Korean Circ J. 2024. PMID: 38956941 Free PMC article. Review.
-
Durable-polymer versus biodegradable-polymer drug-eluting stents in acute coronary syndromes: three-year outcomes of the HOST REDUCE POLYTECH RCT Trial.EuroIntervention. 2024 Jun 17;20(12):e750-e759. doi: 10.4244/EIJ-D-23-01053. EuroIntervention. 2024. PMID: 38887886 Clinical Trial.
Publication types