Overexpansion of the SAPIEN 3 Transcatheter Heart Valve: An Ex Vivo Bench Study
- PMID: 30190060
- DOI: 10.1016/j.jcin.2018.06.027
Overexpansion of the SAPIEN 3 Transcatheter Heart Valve: An Ex Vivo Bench Study
Abstract
Objectives: This study assessed the effect of overexpansion beyond labeled size (diameter) of transcatheter heart valves through an ex vivo bench study.
Background: Transcatheter heart valves function optimally when expanded to specific dimensions. However, clinicians may sometimes wish to overexpand balloon-expandable valves to address specific clinical challenges. The implications of overexpansion have assumed considerable importance, and objective information to guide practice is limited.
Methods: We evaluated SAPIEN 3 transcatheter heart valves (Edwards Lifesciences, Irvine, California). Valves (diameters of 23, 26, and 29 mm) were expanded to nominal dimensions, and then incrementally overexpanded with balloons sized 1-, 2-, and 3-mm larger than the recommended diameter. Valves underwent visual, microcomputed tomography, and hydrodynamic evaluation at various degrees of overexpansion.
Results: SAPIEN 3 valves with labeled diameters of 23, 26, and 29 mm could be incrementally overexpanded to midvalve diameters of 26.4, 28.4, and 31.2 mm, respectively. With overexpansion, there was visible restriction of the valve leaflets, which was particularly evident with the smaller valves. After maximal overexpansion of a 26-mm valve a leaflet tear was observed. High-speed video demonstrated impaired leaflet motion of both the 23- and 26-mm valves and hydrodynamic testing documented a regurgitant fraction for the 23- and 26-mm valves above accepted international standards. The maximally overexpanded 29-mm SAPIEN 3 still had relatively normal leaflet motion and excellent hydrodynamic function. Durability was not specifically evaluated.
Conclusions: Overexpansion of balloon-expandable valves is possible. However, excessive overexpansion may be associated with impaired hydrodynamic function, acute leaflet failure, and reduced durability. Smaller valves may be at greater risk with overexpansion than larger valves. Overexpansion is best avoided unless clinical circumstances are compelling.
Keywords: overexpansion; transcatheter heart valve.
Copyright © 2018 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Comment in
-
Optimizing Valve Sizing in Balloon-Expandable Transcatheter Aortic Valve Replacement.JACC Cardiovasc Interv. 2018 Sep 10;11(17):1706-1709. doi: 10.1016/j.jcin.2018.07.025. JACC Cardiovasc Interv. 2018. PMID: 30190061 No abstract available.
Similar articles
-
Overexpansion of older generation balloon expandable transcatheter heart valves: An ex-vivo bench study.Catheter Cardiovasc Interv. 2019 Nov 15;94(6):806-811. doi: 10.1002/ccd.28160. Epub 2019 Mar 9. Catheter Cardiovasc Interv. 2019. PMID: 30851079
-
Long-Term Durability of Transcatheter Heart Valves: Insights From Bench Testing to 25 Years.JACC Cardiovasc Interv. 2020 Jan 27;13(2):235-249. doi: 10.1016/j.jcin.2019.07.049. Epub 2019 Sep 28. JACC Cardiovasc Interv. 2020. PMID: 31575516
-
Stent and leaflet stresses in a 26-mm first-generation balloon-expandable transcatheter aortic valve.J Thorac Cardiovasc Surg. 2017 May;153(5):1065-1073. doi: 10.1016/j.jtcvs.2016.12.016. Epub 2016 Dec 23. J Thorac Cardiovasc Surg. 2017. PMID: 28108064 Free PMC article.
-
Transcatheter aortic valve replacement using the Edwards SAPIEN transcatheter heart valves.Cardiol Clin. 2013 Aug;31(3):337-50. doi: 10.1016/j.ccl.2013.05.009. Cardiol Clin. 2013. PMID: 23931098 Review.
-
Balloon-expandable prostheses for transcatheter aortic valve replacement.Prog Cardiovasc Dis. 2014 May-Jun;56(6):583-95. doi: 10.1016/j.pcad.2014.02.001. Epub 2014 Mar 1. Prog Cardiovasc Dis. 2014. PMID: 24838134 Review.
Cited by
-
Catheter-based Interventions in Tetralogy of Fallot Across the Lifespan.CJC Pediatr Congenit Heart Dis. 2023 Sep 15;2(6Part A):339-351. doi: 10.1016/j.cjcpc.2023.09.005. eCollection 2023 Dec. CJC Pediatr Congenit Heart Dis. 2023. PMID: 38161670 Free PMC article. Review.
-
A case of a transcatheter valve-in-valve implantation using balloon expandable valve for failed SOLO SMART stentless bioprosthetic valve.J Cardiol Cases. 2023 Jul 1;28(4):168-171. doi: 10.1016/j.jccase.2023.06.006. eCollection 2023 Oct. J Cardiol Cases. 2023. PMID: 37818436 Free PMC article.
-
Calcium Fracture and Device Over Expansion in Transcatheter Aortic Valve Replacement for Bicuspid Aortic Valves.Ann Biomed Eng. 2023 Oct;51(10):2172-2181. doi: 10.1007/s10439-023-03246-6. Epub 2023 May 23. Ann Biomed Eng. 2023. PMID: 37219698
-
Assessment of consistency in aortic valve sizing between surgical and transcatheter aortic valve replacement.Postepy Kardiol Interwencyjnej. 2023 Mar;19(1):40-46. doi: 10.5114/aic.2023.125154. Epub 2023 Feb 16. Postepy Kardiol Interwencyjnej. 2023. PMID: 37090206 Free PMC article.
-
Aortic valve disorders and left ventricular assist devices.Front Cardiovasc Med. 2023 Feb 23;10:1098348. doi: 10.3389/fcvm.2023.1098348. eCollection 2023. Front Cardiovasc Med. 2023. PMID: 36910539 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources