Bimodal Hearing or Bilateral Cochlear Implants? Ask the Patient

Ear Hear. 2019 May/Jun;40(3):501-516. doi: 10.1097/AUD.0000000000000657.

Abstract

Objective: The objectives of this study were to assess the effectiveness of various measures of speech understanding in distinguishing performance differences between adult bimodal and bilateral cochlear implant (CI) recipients and to provide a preliminary evidence-based tool guiding clinical decisions regarding bilateral CI candidacy.

Design: This study used a multiple-baseline, cross-sectional design investigating speech recognition performance for 85 experienced adult CI recipients (49 bimodal, 36 bilateral). Speech recognition was assessed in a standard clinical test environment with a single loudspeaker using the minimum speech test battery for adult CI recipients as well as with an R-SPACE 8-loudspeaker, sound-simulation system. All participants were tested in three listening conditions for each measure including each ear alone as well as in the bilateral/bimodal condition. In addition, we asked each bimodal listener to provide a yes/no answer to the question, "Do you think you need a second CI?"

Results: This study yielded three primary findings: (1) there were no significant differences between bimodal and bilateral CI performance or binaural summation on clinical measures of speech recognition, (2) an adaptive speech recognition task in the R-SPACE system revealed significant differences in performance and binaural summation between bimodal and bilateral CI users, with bilateral CI users achieving significantly better performance and greater summation, and (3) the patient's answer to the question, "Do you think you need a second CI?" held high sensitivity (100% hit rate) for identifying likely bilateral CI candidates and moderately high specificity (77% correct rejection rate) for correctly identifying listeners best suited with a bimodal hearing configuration.

Conclusions: Clinics cannot rely on current clinical measures of speech understanding, with a single loudspeaker, to determine bilateral CI candidacy for adult bimodal listeners nor to accurately document bilateral benefit relative to a previous bimodal hearing configuration. Speech recognition in a complex listening environment, such as R-SPACE, is a sensitive and appropriate measure for determining bilateral CI candidacy and also likely for documenting bilateral benefit relative to a previous bimodal configuration. In the absence of an available R-SPACE system, asking the patient whether or not s/he thinks s/he needs a second CI is a highly sensitive measure, which may prove clinically useful.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Clinical Decision-Making
  • Cochlear Implantation / methods*
  • Cochlear Implants
  • Female
  • Hearing Aids*
  • Hearing Loss, Bilateral / rehabilitation*
  • Hearing Loss, Sensorineural / rehabilitation*
  • Humans
  • Male
  • Middle Aged
  • Patient Reported Outcome Measures
  • Speech Perception*
  • Young Adult