Fistuloplasty using a radiation-and-time-saving sheathless balloon catheter

J Vasc Access. 2019 May;20(3):276-280. doi: 10.1177/1129729818804991. Epub 2018 Oct 14.

Abstract

Introduction: Sheath placement in dialysis access interventions is traditionally necessary to obtain imaging, guide percutaneous angioplasty, and evaluate results. The aim of this study was to assess the feasibility of performing sheathless Arterio-venous (AV) access interventions using a novel percutaneous angioplasty balloon catheter.

Methods: Between May and September 2017, data on all dialysis access interventions using a novel percutaneous angioplasty balloon with a dedicated injection port were collected. All procedures were performed without a sheath. Success was established as no conversion to sheath placement. Demographic data, location of lesion, time to perform procedure, amount of contrast used, radiation exposure, and access complications were recorded. Ultrasound was used to evaluate access site complications.

Results: Sheathless interventions were successful in 24 patients with the mean age of 62 years (29-94). There were 5 PTFE grafts and 19 native fistulas. Lesions were located anywhere from the arterial anastomosis to the cephalic arch. The average balloon size was 6 mm (5-7 mm), and the procedure time was 15.8 min (8-45 min). No access site complications were observed.

Conclusion: Sheathless intervention is feasible with several potential advantages, including short procedure time, minimal contrast volume, and reduced radiation exposure. Finally, the lower profile at the access site may result in fewer complications.

Keywords: Angiography; angioplasty; fistula; phlebography; renal dialysis; ultrasonography.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Angioplasty, Balloon / adverse effects
  • Angioplasty, Balloon / instrumentation*
  • Arteriovenous Shunt, Surgical / adverse effects*
  • Arteriovenous Shunt, Surgical / instrumentation
  • Blood Vessel Prosthesis
  • Blood Vessel Prosthesis Implantation / adverse effects*
  • Blood Vessel Prosthesis Implantation / instrumentation
  • Equipment Design
  • Feasibility Studies
  • Female
  • Graft Occlusion, Vascular / diagnostic imaging
  • Graft Occlusion, Vascular / etiology
  • Graft Occlusion, Vascular / physiopathology
  • Graft Occlusion, Vascular / therapy*
  • Humans
  • Male
  • Middle Aged
  • Polytetrafluoroethylene
  • Prospective Studies
  • Prosthesis Design
  • Radiation Dosage*
  • Radiation Exposure / adverse effects
  • Radiation Exposure / prevention & control*
  • Radiography, Interventional* / adverse effects
  • Renal Dialysis
  • Time Factors
  • Treatment Outcome
  • Vascular Access Devices*
  • Vascular Patency

Substances

  • Polytetrafluoroethylene