Visual and Semiquantitative Accuracy in Clinical Baseline 123I-Ioflupane SPECT/CT Imaging

Clin Nucl Med. 2019 Jan;44(1):1-3. doi: 10.1097/RLU.0000000000002333.

Abstract

Purpose: We aimed to (a) elucidate the concordance of visual assessment of an initial I-ioflupane scan by a human interpreter with comparison to results using a fully automatic semiquantitative method and (b) to assess the accuracy compared to follow-up (f/u) diagnosis established by movement disorder specialists.

Methods: An initial I-ioflupane scan was performed in 382 patients with clinically uncertain Parkinsonian syndrome. An experienced reader performed a visual evaluation of all scans independently. The findings of the visual read were compared with semiquantitative evaluation. In addition, available f/u clinical diagnosis (serving as a reference standard) was compared with results of the human read and the software.

Results: When comparing the semiquantitative method with the visual assessment, discordance could be found in 25 (6.5%) of 382 of the cases for the experienced reader (ĸ = 0.868). The human observer indicated region of interest misalignment as the main reason for discordance. With neurology f/u serving as reference, the results of the reader revealed a slightly higher accuracy rate (87.7%, ĸ = 0.75) compared to semiquantification (86.2%, ĸ = 0.719, P < 0.001, respectively). No significant difference in the diagnostic performance of the visual read versus software-based assessment was found.

Conclusions: In comparison with a fully automatic semiquantitative method in I-ioflupane interpretation, human assessment obtained an almost perfect agreement rate. However, compared to clinical established diagnosis serving as a reference, visual read seemed to be slightly more accurate as a solely software-based quantitative assessment.

Publication types

  • Evaluation Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Female
  • Humans
  • Image Interpretation, Computer-Assisted / methods*
  • Image Interpretation, Computer-Assisted / standards
  • Male
  • Middle Aged
  • Nortropanes
  • Parkinsonian Disorders / diagnostic imaging*
  • Radiopharmaceuticals
  • Single Photon Emission Computed Tomography Computed Tomography / methods
  • Single Photon Emission Computed Tomography Computed Tomography / standards*

Substances

  • Nortropanes
  • Radiopharmaceuticals
  • ioflupane