Fatigue and microgap behaviour of a three-unit implant-fixed dental prosthesis combining conventional and dynamic abutments
- PMID: 30422735
- DOI: 10.1080/03091902.2018.1520312
Fatigue and microgap behaviour of a three-unit implant-fixed dental prosthesis combining conventional and dynamic abutments
Abstract
This is an in vitro study composed by a fatigue test followed by an optical microscopy analysis. Dynamic abutments concept, recently introduced on screw-retained implant dental prosthesis, consists on the screw channel customisation according to the individual needs of each rehabilitation. Geometry and tightening torque differences advise the assessment of their mechanical performance. Clarify whether the combination of dynamic and conventional abutments in a three-unit implant-fixed prosthesis has detrimental effects either on the mechanical performance under cyclic loading or on the implant-abutment microgap dimensions. The fatigue test was performed in agreement with the ISO standard 14801. Then on the samples that resisted 5 million cycles, the implant-abutment microgap was measured on dynamic and conventional abutments using optical microscopy. Two unloaded samples were used as control group. The samples supported a load of 1050 N. The implant-abutment microgap measurement did not show statistically significant differences (p = .086) between loaded and unloaded groups, but the loaded conventional abutments showed a significant lower implant-abutment microgap (p = .05) than the loaded dynamic abutments. The combination of conventional and dynamic abutments do not seem to produce a decrease in fatigue resistance to a level below the mastication forces or an increase in the joint dimensions.
Keywords: Dynamic abutments; fatigue test; microgap.
Similar articles
-
Removal Torque and Biofilm Accumulation at Two Dental Implant-Abutment Joints After Fatigue.Int J Oral Maxillofac Implants. 2016 Jul-Aug;31(4):813-9. doi: 10.11607/jomi.4173. Int J Oral Maxillofac Implants. 2016. PMID: 27447147
-
Effect of cyclic loading on the vertical microgap of long-span zirconia frameworks supported by 4 or 6 implants.J Prosthet Dent. 2014 Oct;112(4):828-33. doi: 10.1016/j.prosdent.2014.03.007. Epub 2014 Apr 29. J Prosthet Dent. 2014. PMID: 24795260
-
Evaluation of the sealing capability of implants to titanium and zirconia abutments against Porphyromonas gingivalis, Prevotella intermedia, and Fusobacterium nucleatum under different screw torque values.J Prosthet Dent. 2014 Sep;112(3):561-7. doi: 10.1016/j.prosdent.2013.11.010. Epub 2014 Mar 20. J Prosthet Dent. 2014. PMID: 24656409
-
Fracture strength of implant abutments after fatigue testing: A systematic review and a meta-analysis.J Mech Behav Biomed Mater. 2016 Sep;62:333-346. doi: 10.1016/j.jmbbm.2016.05.011. Epub 2016 May 14. J Mech Behav Biomed Mater. 2016. PMID: 27239815 Review.
-
[Advance and review: implant-abutment microgap].Zhonghua Kou Qiang Yi Xue Za Zhi. 2017 Nov 9;52(11):704-708. doi: 10.3760/cma.j.issn.1002-0098.2017.11.012. Zhonghua Kou Qiang Yi Xue Za Zhi. 2017. PMID: 29972953 Review. Chinese.
Cited by
-
The effect of mandibular flexure on the design of implant-supported fixed restorations of different facial types under two loading conditions by three-dimensional finite element analysis.Front Bioeng Biotechnol. 2022 Aug 29;10:928656. doi: 10.3389/fbioe.2022.928656. eCollection 2022. Front Bioeng Biotechnol. 2022. PMID: 36105608 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources