Assessment of four different cardiac softwares for evaluation of LVEF with CZT-SPECT vs CMR in 48 patients with recent STEMI

J Nucl Cardiol. 2020 Dec;27(6):2017-2026. doi: 10.1007/s12350-018-01493-y. Epub 2018 Nov 13.

Abstract

Purpose: To compare, vs CMR, four softwares: quantitative gated SPECT (QGS), myometrix (MX), corridor 4DM (4DM), and Emory toolbox (ECTb) to evaluate left ventricular ejection fraction (LVEF), end-systolic (ESV), and end-diastolic volumes (EDVs) by gated MPI CZT-SPECT.

Methods: 48 patients underwent MPI CZT-SPECT and CMR 6 weeks after STEMI, LV parameters were measured with four softwares at MPI CZT-SPECT vs CMR. We evaluated (i) concordance and correlation between MPI CZT-SPECT and CMR, (ii) concordance MPI CZT-SPECT/CMR for the categorical evaluation of the left ventricular dysfunction, and (iii) impacts of perfusion defects > 3 segments on concordance.

Results: LVEF: LCC QGS/CMR = 0.81 [+ 2.2% (± 18%)], LCC MX/CMR = 0.83 [+ 1% (± 17.5%)], LCC 4DM/CMR = 0.73 [+ 3.9% (± 21%)], LCC ECTb/CMR = 0.69 [+ 6.6% (± 21.1%)]. ESV: LCC QGS/CMR = 0.90 [- 8 mL (± 40 mL)], LCC MX/CMR = 0.90 [- 9 mL (± 36 mL)], LCC 4DM/CMR = 0.89 [+ 4 mL (± 45 mL)], LCC ECTb/CMR = 0.87 [- 3 mL (± 45 mL)]. EDV: LCC QGS/CMR = 0.70 [- 16 mL (± 67 mL)], LCC MX/CMR = 0.68 [- 21 mL (± 63 mL], LCC 4DM/CMR = 0.72 [+ 9 mL (± 73 mL)], LCC ECTb/CMR = 0.69 [+ 10 mL (± 70 mL)].

Conclusion: QGS and MX were the two best-performing softwares to evaluate LVEF after recent STEMI.

Keywords: Left ventricular ejection fraction; MPI CZT-SPECT; end-diastolic volume; end-systolic volume; softwares.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Cadmium*
  • Diastole
  • Female
  • Gated Blood-Pool Imaging / methods*
  • Humans
  • Image Processing, Computer-Assisted / instrumentation*
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Middle Aged
  • ST Elevation Myocardial Infarction / diagnostic imaging*
  • Software
  • Stroke Volume
  • Technetium Tc 99m Sestamibi
  • Tellurium*
  • Tomography, Emission-Computed, Single-Photon / methods*
  • Ventricular Dysfunction, Left
  • Ventricular Function, Left*
  • Zinc*

Substances

  • CdZnTe
  • Cadmium
  • Technetium Tc 99m Sestamibi
  • Zinc