Newborn screening for 3-hydroxy-3-methylglutaric aciduria using direct analysis in real-time mass spectrometry

J Mass Spectrom. 2019 Feb;54(2):134-140. doi: 10.1002/jms.4314.

Abstract

A novel method utilizing ambient thermal desorption ionization with a direct analysis in real-time source integrated with mass spectrometry (DART-MS) was established and applied to the rapid analysis of 3-hydroxy-3-methylglutaric (3-HMG) acid in the neonatal urine. Instrument parameter settings were optimized to obtain high sensitive and accurate determination of 3-HMG acid. The use of helium gas heated to temperature of 400°C was observed to permit deprotonation, 3-HMG acid producing an abundant (M-H)- (m/z 161) in the negative ion mode. The calibration curve was determined to be linear over the range of 0.05-5 mg/L, with the correlation coefficient r = 0.9988 and the relative standard deviations (n = 6) in the range of 1.5-11.8%. The limit of detection was 0.002 mg/L, and the limit of quantitation was 0.007 mg/L. The recoveries ranged from 88.0% to 123.1%. Four urine samples from patients and four simulated urine samples were investigated. The results of DART-MS were in agreement with the values determined using established methods at the hospitals. The proposed method demonstrated significant potential in the application of the high-throughput screening in newborn screening.

Keywords: 3-Hydroxy-3-methylglutaric acid; direct analysis in real-time mass spectrometry; newborn screening; urine.

MeSH terms

  • Acetyl-CoA C-Acetyltransferase / deficiency*
  • Acetyl-CoA C-Acetyltransferase / urine
  • Amino Acid Metabolism, Inborn Errors / diagnosis*
  • Amino Acid Metabolism, Inborn Errors / urine
  • Humans
  • Infant, Newborn
  • Mass Spectrometry / methods
  • Neonatal Screening

Substances

  • Acetyl-CoA C-Acetyltransferase

Supplementary concepts

  • 3-Hydroxy-3-Methylglutaryl-CoA Lyase Deficiency