More reoperations for periprosthetic fracture after cemented hemiarthroplasty with polished taper-slip stems than after anatomical and straight stems in the treatment of hip fractures: a study from the Norwegian Hip Fracture Register 2005 to 2016

Bone Joint J. 2018 Dec;100-B(12):1565-1571. doi: 10.1302/0301-620X.100B12.BJJ-2018-0262.R1.

Abstract

Aims: The aim of this large registry-based study was to compare mid-term survival rates of cemented femoral stems of different designs used in hemiarthroplasty for a fracture of the femoral neck.

Patients and methods: From the Norwegian Hip Fracture Register (NHFR), 20 532 primary cemented bipolar hemiarthroplasties, which were undertaken in patients aged > 70 years with a femoral neck fracture between 2005 and 2016, were included. Polished tapered stems (n = 12 065) (Exeter and CPT), straight stems (n = 5545) (Charnley, Charnley Modular, and Spectron EF), and anatomical stems (n = 2922) (Lubinus SP2) were included. The survival of the implant with any reoperation as the endpoint was calculated using the Kaplan-Meier method and hazard ratios (HRs), and the different indications for reoperation were calculated using Cox regression analysis.

Results: The one-year survival was 96.0% (95% confidence interval (CI) 95.6 to 96.4) for the Exeter stem, 97.0% (95% CI 96.4 to 97.6) for the Lubinus SP2 stem, 97.6% (95% CI 97.0 to 98.2) for the Charnley stem, 98.1% (95% CI 97.3 to 98.9) for the Spectron EF stem, and 96.4% (95% CI 95.6 to 97.2) for the Charnley Modular stem, respectively. The hazard ratio for reoperation after one year was lower for Lubinus SP2 (HR 0.77, 95% CI 0.60 to 0.97), Charnley (HR 0.64, 95% CI 0.48 to 0.86), and Spectron EF stems (HR 0.44, 95% CI 0.29 to 0.67) compared with the Exeter stem. Reoperation for periprosthetic fracture occurred almost exclusively after the use of polished tapered stems.

Conclusion: We were able to confirm that implant survival after cemented hemiarthroplasty for a hip fracture is high. Differences in rates of reoperation seem to favour anatomical and straight stems compared with polished tapered stems, which had a higher risk of periprosthetic fracture.

Keywords: Hemiarthroplasty; Periprosthetic fracture; Reoperation.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Bone Cements*
  • Female
  • Follow-Up Studies
  • Hemiarthroplasty / adverse effects*
  • Hip Fractures / diagnosis
  • Hip Fractures / surgery*
  • Hip Prosthesis
  • Humans
  • Incidence
  • Male
  • Norway / epidemiology
  • Periprosthetic Fractures / epidemiology
  • Periprosthetic Fractures / etiology
  • Periprosthetic Fractures / surgery*
  • Postoperative Complications / epidemiology
  • Postoperative Complications / etiology
  • Postoperative Complications / surgery*
  • Prospective Studies
  • Prosthesis Design
  • Registries*
  • Reoperation / statistics & numerical data*
  • Risk Factors
  • Treatment Outcome

Substances

  • Bone Cements