Accuracy, Utilization, and Effectiveness Comparisons of Different Continuous Glucose Monitoring Systems

Diabetes Technol Ther. 2019 Mar;21(3):128-132. doi: 10.1089/dia.2018.0374. Epub 2019 Jan 25.

Abstract

Background: Accuracy and feature sets of continuous glucose monitoring (CGM) systems may influence device utilization and outcomes. We compared clinical trial accuracy and real-world utilization and effectiveness of two different CGM systems.

Materials and methods: Separately conducted accuracy studies of a fifth-generation and a sixth-generation CGM system involved 50 and 159 adults, respectively. For between-system performance comparisons, propensity score methods were utilized to balance cohort characteristics. Real-world outcomes were assessed in 10,000 anonymized patients who had switched from the fifth-generation to the sixth-generation system and had used connected mobile devices to upload data from both systems, allowing pairwise comparisons of device utilization and glucose concentration distributions.

Results: Propensity score-adjusted mean absolute relative differences for the fifth- and sixth-generation systems were 9.0% and 9.9%, and the percentages of values within ±20%/20 mg/dL were 93.1% and 92.5%, respectively. The sixth-generation system, but not the fifth-generation system, met accuracy criteria for interoperable CGM systems. Both systems had high real-world utilization rates (93.8% and 95.3% in the fifth- and sixth-generation systems, respectively). Use of the sixth-generation system was associated with fewer glucose values <55 mg/dL (<3.1 mmol/L) (0.7% vs. 1.1%, P < 0.001) and more values 70-180 mg/dL (3.9-10.0 mmol/L) (57.3% vs. 56.0%, P < 0.001) than the fifth-generation system.

Conclusions: CGM performance outcomes can be compared through the propensity score analysis of clinical trial data and pairwise comparisons of real-world data. The systems compared here had nearly equivalent accuracy and utilization rates. Longer term biochemical and psychosocial benefits observed with the fifth-generation system are also expected with the sixth-generation system.

Keywords: Continuous glucose monitoring; Dexcom G5; Dexcom G6; Propensity scoring; Real-world evidence..

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Blood Glucose / analysis*
  • Blood Glucose Self-Monitoring / instrumentation*
  • Blood Glucose Self-Monitoring / psychology*
  • Diabetes Mellitus / blood
  • Diabetes Mellitus / psychology*
  • Female
  • Humans
  • Male
  • Patient Acceptance of Health Care / psychology*
  • Propensity Score

Substances

  • Blood Glucose