Compressed SENSE single-breath-hold and free-breathing cine imaging for accelerated clinical evaluation of the left ventricle

Clin Radiol. 2019 Apr;74(4):325.e9-325.e17. doi: 10.1016/j.crad.2018.12.012. Epub 2019 Jan 25.

Abstract

Aim: To assess the accuracy of compressed SENSE (CS-SENSE) cine cardiac magnetic resonance imaging (CMR) with and without breath-hold in comparison to standard cine CMR with breath-hold for the assessment of left ventricular (LV) function.

Materials and methods: Thirty-three healthy volunteers underwent balanced turbo field-echo cine CMR with breath-hold (BTFE-BH; reference standard), single breath-hold CS-SENSE (csBTFE-BH) cine CMR, and free-breathing (FB) CS-SENSE (csBTFE-FB) cine CMR on a 3 T MRI system. All images were acquired in stacks of eight short-axis sections. Image quality was assessed and compared by the Wilcoxon matched-pairs signed-rank test. End-diastolic volume, end-systolic volume, stroke volume, ejection fraction, LV end-diastolic (LVED) mass, regional myocardial wall motion, and scan time were compared by paired t-test, linear regression, and Bland-Altman analyses.

Results: All techniques provided acceptable image quality (score ≥3) for LV volumetric analysis in all participants (BTFE-BH [reference standard]: 5.00±0.00; csBTFE-BH: 4.03±0.17 [p<0.001]; csBTFE-FB: 3.76±0.44 [p<0.001]), with good agreement in LV function assessment; however, there was a slight but significant underestimation of LVED mass by csBTFE-FB (csBTFE-FB: 73.63±17.31 g versus BTFE-BH [reference standard]: 75.12±18.18 g, p=0.037). All methods showed a strong correlation with quantitative regional myocardial wall motion. Acquisition times for both csBTFE-BH and csBTFE-FB were significantly shorter than that for BTFE-BH (BTFE-BH [reference standard]: 89.3±5.70 seconds; csBTFE-BH: 24.42±2.18 seconds [p<0.001]; csBTFE-FB: 22.48±1.85 seconds [p<0.001]).

Conclusion: LV function assessment with the novel CS-SENSE cine CMR is not inferior to standard cine CMR, irrespective of BH; however, LVED mass is underestimated by csBTFE-FB.

MeSH terms

  • Adult
  • Breath Holding*
  • Female
  • Heart Ventricles / diagnostic imaging*
  • Humans
  • Image Interpretation, Computer-Assisted
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging, Cine / methods*
  • Male
  • Prospective Studies
  • Reference Values
  • Reproducibility of Results
  • Respiration*