Retrieval analysis of contemporary antioxidant polyethylene: multiple material and design changes may decrease implant performance

Knee Surg Sports Traumatol Arthrosc. 2019 Jul;27(7):2111-2119. doi: 10.1007/s00167-019-05387-5. Epub 2019 Feb 10.

Abstract

Purpose: With the introduction of the Attune Knee System (DePuy) in March 2013, a new polyethylene formulation incorporating anti-oxidants was used. Although several in vitro studies have demonstrated the positive effects of antioxidants on UHMWPE, no retrieval study has looked at polyethylene damage of this system yet. It was the aim of this study to investigate the in vivo performance of this new design, by comparing it with its predecessors in retrieval analysis.

Methods: 24 PFC (18 fixed bearing and 6 rotating platform designs) and 17 Attune (8 fixed bearing and 9 rotating platform designs) implants were retrieved. For retrieval analysis, a macroscopic analysis of polyethylene components, using a peer-reviewed damage grading method was used. Medio-lateral polyethylene thickness difference was measured with a peer-reviewed micro-CT based method. The roughness of metal components was measured. All findings were compared between the two designs.

Results: Attune tibial inserts with fixed bearings showed significantly higher hood scores on the backside surface when compared with their PFC counterparts (p = 0.01), no other significant differences were found in the polyethylene damage of all the other surfaces analysed, in the surface roughness of metal components and in medio-lateral linear deformations.

Conclusion: A significant difference between PFC and Attune fixed bearing designs was found in terms of backside surface damage: multiple changes in material and design features could lead to a potential decrease of implant performance. Results from the present study may help to understand how the new Attune Knee System performs in vivo, impacting over 600,000 patients.

Keywords: Polyethylene; Polyethylene linear deformation; Polyethylene surface damage; Retrieval analysis; Total knee arthroplasty.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Antioxidants*
  • Arthroplasty, Replacement, Knee / instrumentation*
  • Female
  • Humans
  • Knee Joint / surgery
  • Knee Prosthesis*
  • Male
  • Metals
  • Middle Aged
  • Polyethylene*
  • Prosthesis Design
  • Prosthesis Failure
  • Stress, Mechanical
  • X-Ray Microtomography