Context.—: Infectious gastroenteritis is caused by various pathogens, including bacteria, viruses, and parasites.
Objective.—: To compare the performance of Seegene Allplex Gastrointestinal (24 targets: 13 bacteria, 5 viruses, and 6 parasites in 4 panels), Luminex xTAG Gastrointestinal Pathogen Panel (15 targets: 9 bacteria, 3 viruses, and 3 parasites), and BD MAX Enteric panel (5 bacteria and 3 parasites). We estimated the agreement among 3 molecular assays.
Design.—: A total of 858 stool samples (554 bacterial/parasite and 304 viral pathogens) were included. A consensus positive/negative was defined as concordant results from at least 2 tests. To evaluate the agreement among the assays, κ value was calculated.
Results.—: The overall positive percentage agreements of Seegene, Luminex, and BD MAX were 94% (258 of 275), 92% (254 of 275), and 78% (46 of 59), respectfully. For Salmonella, Luminex showed low negative percentage agreement because of frequent false positives (n = 31) showing low median fluorescent intensity. For viruses, positive/negative percentage agreements of Seegene and Luminex were 99%/96% and 93%/99%, respectively. Compared with routine microbiology testing, Seegene, Luminex, and BD MAX additionally identified 39, 40, and 12 pathogens, respectively. Sixty-one cases (16 cases with Seegene, 51 cases with Luminex, and 1 case with BD MAX) showed positive results for multiple pathogens, but only 3 were consensus positive.