Fricke xylenol glycine gel dosimeter for in vivo dosimetry at extended treatment distance

Appl Radiat Isot. 2019 Mar:145:217-222. doi: 10.1016/j.apradiso.2018.09.028. Epub 2018 Sep 27.

Abstract

Fricke xylenol gelatin with glycine (FXGGly) as gel dosimeter was investigated for in vivo dosimetry at extended source-to-surface distance (eSSD) and validated with optically stimulated luminescence dosimeters (OSLD). FXGGly dosimeter was calibrated with a low dose rate of 10 cGy/min at eSSD and evaluated using anthropomorphic phantom simulating the actual treatment scenario. The FXGGly dosimeter was demonstrated as a suitable in vivo dosimeter for total body irradiation at eSSD with desirable dosimetric accuracy (± 2%) in comparison with OSLD or ionization chamber.