Comparison of the clinical performance of i-gel and Ambu AuraGain in children: A randomised noninferiority clinical trial

Eur J Anaesthesiol. 2019 Jun;36(6):411-417. doi: 10.1097/EJA.0000000000000987.

Abstract

Background: Supraglottic airway devices are commonly used in general anaesthesia in children. The AuraGain is a newly developed disposable second-generation supraglottic airway device with an inflatable cuff and a gastric port.

Objective: To confirm our hypothesis that the oropharyngeal leak pressure (OLP) of the AuraGain would be noninferior to that of the i-gel in paediatric patients.

Design: A single-blinded, randomised, noninferiority, clinical trial.

Setting: Single-centre trial from January to March 2017.

Patients: One hundred paediatric patients (American Society of Anesthesiologists' physical status 1 to 3), aged up to 12 years old, and body weight of 5 to 30 kg requiring a supraglottic airway for elective surgery with an expected surgery time of less than 2 h under general anaesthesia.

Intervention: The patients were randomised to allocation to the AuraGain group or the i-gel group. The device size (1.5 to 2.5) used in each group was based on the manufacturer's recommendation.

Main outcome measures: The primary outcome measure was OLP immediately after insertion, with a predefined noninferiority margin of 3 cmH2O.

Results: The OLP immediately after insertion was lower for the AuraGain than for the i-gel (17.1 vs. 23.0 cmH2O; mean difference: -5.9 cmH2O; 95% confidence interval: -8.5 to -3.3 cmH2O; P = 0.98 and <0.001 for noninferiority and superiority, respectively). The first-attempt success rate (AuraGain, 96% vs. i-gel, 90%; P = 0.44) was comparable between the devices. The incidence of blood staining was lower (AuraGain, 6% vs. i-gel, 0%; P = 0.012) and time to insertion was shorter (AuraGain 21.3 vs. i-gel, 17.1 s; P < 0.001) for the i-gel compared with the AuraGain.

Conclusion: Our noninferiority hypothesis was not adopted. The OLP of the i-gel may be higher than that of the AuraGain, but this superiority hypothesis should be investigated in future trials. The i-gel demonstrated a shorter time to successful placement and lower incidence of blood staining than the AuraGain.

Trial registration: University Hospital Medical Information Network identifier, UMIN000024222.

Publication types

  • Equivalence Trial
  • Randomized Controlled Trial

MeSH terms

  • Anesthesia, General / adverse effects
  • Anesthesia, General / instrumentation*
  • Anesthesia, General / methods
  • Child, Preschool
  • Disposable Equipment
  • Elective Surgical Procedures / adverse effects*
  • Equipment Design
  • Female
  • Hemorrhage / epidemiology
  • Hemorrhage / etiology
  • Humans
  • Incidence
  • Infant
  • Intraoperative Complications / epidemiology*
  • Intraoperative Complications / etiology
  • Laryngeal Masks / adverse effects*
  • Male
  • Oropharynx / injuries*
  • Pharyngeal Diseases / epidemiology
  • Pharyngeal Diseases / etiology
  • Pressure