Optimization of CRT programming using non-invasive electrocardiographic imaging to assess the acute electrical effects of multipoint pacing

J Arrhythm. 2019 Jan 14;35(2):267-275. doi: 10.1002/joa3.12153. eCollection 2019 Apr.

Abstract

Aim: Quadripolar lead technology and multi-point pacing (MPP) are important clinical adjuncts in cardiac resynchronization therapy (CRT) pacing aimed at reducing the rate of non-response to therapy. Mixed results have been achieved using MPP and it is critical to identify which patients require this approach and how to configure their MPP stimulation, in order to achieve optimal electrical resynchronization.

Methods & results: We sought to investigate whether electrocardiographic imaging (ECGi), using the CARDIOINSIGHT inverse ECG mapping system, could identify alterations in electrical resynchronization during different methods of device optimization. In no patient did a single form of programming optimization provide the best electrical response. The effects of utilizing MPP were idiosyncratic and highly patient specific. ECGi activation maps were clearly able to discern changes in bulk LV activation during differing MPP programming. In two of the five subjects, MPP resulted in more rapid activation of the left ventricle compared to standard CRT; however, in the remaining three patients, the use of MPP did not appear to acutely improve electrical resynchronization. Crucially, this cohort showed evidence of extensive LV scarring which was well visualized using both CMR and ECGi voltage mapping.

Conclusions: Our work suggests a potential role for ECGi in the optimization of non-responders to CRT, as it allows the fusion of activation maps and scar analysis above and beyond interrogation of the 12 lead ECG.

Keywords: CRT; electrocardiographic mapping; multi‐point pacing; multi‐site pacing; non‐responders.