Design optimization of a pixel-based range telescope for proton computed tomography

Phys Med. 2019 Jul:63:87-97. doi: 10.1016/j.ejmp.2019.05.026. Epub 2019 Jun 6.

Abstract

Purpose: A pixel-based range telescope for tracking particles during proton imaging is described. The detector applies a 3D matrix of stacked Monolithic Active Pixel Sensors with fast readout speeds. This study evaluates different design alternatives of the range telescope on basis of the protons' range accuracy and the track reconstruction efficiency.

Method: Detector designs with different thicknesses of the energy-absorbing plates between each sensor layer are simulated using the GATE/Geant4 Monte Carlo software. Proton tracks traversing the detector layers are individually reconstructed, and a Bragg curve fitting procedure is applied for the calculation of each proton's range.

Results: Simulations show that the setups with 4 mm and thinner absorber layers of aluminum have a low range uncertainty compared to the physical range straggling, systematic errors below 0.3 mm water equivalent thickness and a track reconstruction capability exceeding ten million protons per second.

Conclusions: In order to restrict the total number of layers and to yield the required tracking and range resolution properties, a design recommendation is reached where the proposed range telescope applies 3.5 mm thick aluminum absorber slabs between each sensor layer.

Keywords: Detector optimization; Monte Carlo simulation; Proton computed tomography; Track reconstruction.

MeSH terms

  • Equipment Design
  • Image Processing, Computer-Assisted
  • Monte Carlo Method
  • Phantoms, Imaging
  • Protons*
  • Scattering, Radiation
  • Software
  • Telescopes*
  • Tomography, X-Ray Computed / instrumentation*