On light sharing TOF-PET modules with depth of interaction and 157 ps FWHM coincidence time resolution

Phys Med Biol. 2019 Aug 7;64(15):155008. doi: 10.1088/1361-6560/ab2cb0.

Abstract

The performance of a light sharing and recirculation mechanism that allows the extraction of depth of interaction (DOI) are investigated in this paper, with a particular focus on timing. In parallel, a method to optimize the coincidence time resolution (CTR) of PET detectors by use of the DOI information is proposed and tested. For these purposes, a dedicated 64-channels readout setup has been developed with intrinsic timing resolution of 16 ps FWHM. Several PET modules have been produced, based on LYSO:Ce scintillators and commercial silicon photomultiplier (SiPM) arrays, with [Formula: see text] mm2 individual SiPM size. The results show the possibility to achieve a timing resolution of 157 ps FWHM, combined with the already demonstrated spatial resolution of 1.5 mm FWHM, DOI resolution of 3 mm FWHM, and energy resolution of 9% FWHM at 511 keV, with 15 mm long crystals of section [Formula: see text] mm2 and [Formula: see text] mm2. At the same time, the extraction of the DOI coordinate has been demonstrated not to deteriorate the timing performance of the PET module.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Photons
  • Positron-Emission Tomography / instrumentation
  • Positron-Emission Tomography / methods*
  • Positron-Emission Tomography / standards