Multicenter prospective observational long-term follow-up study of endocardial cardiac resynchronization therapy using the Jurdham procedure

Heart Rhythm. 2019 Oct;16(10):1453-1461. doi: 10.1016/j.hrthm.2019.07.017. Epub 2019 Jul 16.

Abstract

Background: Endocardial cardiac resynchronization therapy (eCRT) avoids the limitations and failures of coronary sinus (CS) resynchronization. However, data regarding long-term outcomes are lacking.

Objective: The purpose of this study was to report the long-term outcome of eCRT performed using the Jurdham procedure in a real-world setting.

Methods: eCRT was performed in patients who failed a CS implant or failed to respond to cardiac resynchronization therapy (CRT), or in selected patients requiring lifelong oral anticoagulation (OAC). Left ventricular ejection fraction (LVEF), New York Heart Association functional class (NYHA FC), and left ventricular stimulation parameters were assessed during long-term follow-up (FU).

Results: From August 2009 to March 2018, the Jurdham procedure was performed in 88 patients at 15 centers in 8 countries, with FU of 32.88 ± 61.52 months (range 0-88 months; 196 patient-years). NYHA FC improved from 2.9 preimplant to 1.3 during FU. LVEF increased <10 percentage points from baseline in 7% of patients, between 10 and 20 percentage points in 11% of patients, and >20 percentage points in 82% of patients. All-cause mortality at 60 months was 30.5%. Three transient ischemic attacks (1.53 per 100 patient-years) and 6 strokes (3.06 per 100 patient-years) occurred. Of the 6 patients with stroke, 4 (66%) had almost complete recovery.

Conclusion: eCRT using the Jurdham procedure is an effective and safe technique in anticoagulated patients. This approach may be an attractive option for patients with failed CS implants or nonresponders to CS CRT. In addition, it might be a reasonable approach as a first option for treatment of patients requiring lifelong OAC.

Keywords: Cardiac resynchronization therapy nonresponders; Endocardial cardiac resynchronization therapy; Endocardial left ventricular lead; Jurdham procedure; Real-life assessment.

Publication types

  • Multicenter Study
  • Observational Study

MeSH terms

  • Aged
  • Cardiac Resynchronization Therapy / methods*
  • Cardiac Resynchronization Therapy / mortality
  • Cardiac Resynchronization Therapy Devices
  • Cause of Death
  • Cohort Studies
  • Female
  • Heart Failure / diagnostic imaging
  • Heart Failure / mortality*
  • Heart Failure / therapy*
  • Heart Function Tests
  • Humans
  • Male
  • Middle Aged
  • Patient Safety / statistics & numerical data
  • Patient Selection
  • Prognosis
  • Retrospective Studies
  • Risk Assessment
  • Severity of Illness Index
  • Stroke Volume / physiology*
  • Survival Rate
  • Treatment Outcome
  • Ventricular Remodeling / physiology*