Here we validate a GC, Flame Ionization Detection (GC-FID), liquid injection method using hydrogen as a carrier gas combining analysis of toxic volatile alcohols (VA): methanol, ethanol, isopropanol, acetone, as well as glycols, ethylene glycol (EG) and propylene glycol (PG), in a single method.
Methodology: 200 μL of calibrator, QC, or patient specimen were deproteinized with 400 μL of acetonitrile containing internal standards (10 mmol/L N-propyl alcohol for VA and 2.5 mmol/L 1,2-butanediol for glycols). GC-FID analysis using hydrogen carrier gas and nitrogen makeup gas utilized an Agilent 7890 system equipped with Agilent 7683 liquid autosampler on a 30 m × 530 μm RTX-200 fused silica column. Method validation included repeatability, recovery, carryover, linearity, lower limit of quantification (LLOQ), accuracy, selectivity and measurement uncertainty.
Results: The 8.3 min from injection to injection reduced time of analysis by 45% over a previously reported method using Helium carrier gas with no loss in resolution. Within-run and Between-run variability were ≤1.4% and ≤6.8% respectively. Recovery was 100% within a 95% confidence interval. Carryover was negligible for all but EG. LLOQ was <1 mmol/L for all analytes. The upper range of linearity was 120 mmol/L for methanol, ethanol and isopropanol, 100 mmol/L for acetone and 50 mmol/L for EG. Analytes demonstrated acceptable accuracy and measurement uncertainty using College of American Pathologists (CAP) criteria. Toluene can cause a false positive EG, while benzene, xylene and 1,3 butanediol can cause false negative EG.
Conclusions: Converting from Helium to Hydrogen carrier gas benefits patient care through a reduction in turnaround time and provides a cost savings to the laboratory.
Copyright © 2019 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.