Resistance to Cyclic Fatigue of Reciprocating Instruments Determined at Body Temperature and Phase Transformation Analysis

Affiliations

Resistance to Cyclic Fatigue of Reciprocating Instruments Determined at Body Temperature and Phase Transformation Analysis

Raymond Scott et al. Aust Endod J. .

Abstract

The purpose of this study was to compare cyclic fatigue (CF) resistance of reciprocating instruments at body temperature and relate the findings to their martensitic transformation temperatures. Contemporary nickel-titanium (NiTi) reciprocating instruments WaveOne Primary, WaveOne Gold Primary and EdgeFile X1 (n = 20 each and #25 tip diameter) were tested for CF resistance at body temperature (37 ± 1°C). Instruments were actioned according to manufacturer guidelines until fracture occurred in a simulated canal (angle of curvature = 60°, radius of curvature = 3 mm and centre of curvature = 5 mm from the tip). Time to fracture was recorded, and data analysed using Weibull analysis. Two instruments of each were tested using differential scanning calorimetry (DSC) to assess phase transformation temperatures. Reciprocating instruments manufactured with new alloys seem to be safer to CF than those manufactured with traditional M-Wire at body temperature. Martensitic transformation temperatures seem not to relate with fatigue behaviour for reciprocating motions.

Keywords: body temperature; cyclic fatigue resistance; differential scanning calorimetry; martensitic transformation temperature.

Similar articles

See all similar articles

References

    1. Peters OA, Bahia MG, Pereira ES. Contemporary root canal preparation: innovations in biomechanics. Dent Clin North Am 2017; 61: 1193-7.
    1. Zupanc J, Vahdat-Pajouh N, Schäfer E. New thermomechanically treated NiTi alloys - a review. Int Endod J 2018; 51: 1088-103.
    1. Bird DC, Chambers D, Peters OA. Usage parameters of nickel-titanium rotary instruments: a survey of endodontists in the United States. J Endod 2009; 35: 1193-7.
    1. Sattapan B, Nervo DJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod 2000; 26: 161-5.
    1. Shen Y, Cheung G, Peng B, Haapasalo M. Defects in nickel-titanium instruments after clinical use. Part 2. Fractographic analysis of fractured surface in cohort study. J Endod 2009; 35: 133-6.

LinkOut - more resources

Feedback