Disentangling time-focusing from beam divergence: A novel approach for high-flux thermal neutron spectroscopy at continuous and long-pulse sources

Rev Sci Instrum. 2019 Sep;90(9):095101. doi: 10.1063/1.5098899.

Abstract

We present the concept of a novel time-focusing technique for neutron spectrometers, which allows us to disentangle time-focusing from beam divergence. The core of this approach is a double rotating-crystal monochromator that can be used to extract a larger wavelength band from a white beam, thus providing a higher flux at the sample compared to standard time-of-flight instruments, yet preserving energy resolution and beam collimation. The performances of a spectrometer based on this approach are quantitatively discussed in terms of possible incident wavelengths, flux at the sample, and (Q, E)-resolution. Analytical estimates suggest flux gains of about one order of magnitude at comparable resolutions in comparison to conventional time-of-flight spectrometers. Moreover, the double monochromator configuration natively shifts the sample away from the source line-of-sight, thus significantly improving the signal-to-noise ratio. The latter, in combination with a system that does not increase the beam divergence, brings the further advantage of a cleaner access to the low-Q region, which is recognized to be of fundamental interest for magnetism and for disordered materials, from glasses to biological systems.