The motor unit and quantitative electromyography

Muscle Nerve. 2020 Feb;61(2):131-142. doi: 10.1002/mus.26718. Epub 2019 Oct 23.

Abstract

Electromyography (EMG) assesses the anatomic motor unit (A-MU), but knowledge of its anatomy, physiology, and changes with pathology is limited. The electrophysiological motor unit (E-MU) and its motor unit potential (E-MUP) represents a fraction of the A-MU. Routine EMG assesses a limited number of E-MUP waveform characteristics (metrics) and their magnitudes qualitatively scaled in a nonlinear manner. Another approach is quantitative EMG (QEMG), whereby 20+ E-MUPs are extracted and both basic and derived metrics obtained and values expressed quantitatively. In diseased muscle, many E-MUP metrics may be normal, which complicates diagnostic interpretation. In QEMG, E-MUP metrics can be clustered and statistical analyses performed to assign probabilities that E-MUPs (and the muscle) are normal, neuropathic, or myopathic. In this article we review what is known about the A-MU, the restricted E-MU, E-MUP metrics, and what QEMG offers currently and in the future.

Keywords: EMG; QEMG; electrodiagnosis; motor unit; motor unit potential; quantitative EMG.

Publication types

  • Review

MeSH terms

  • Action Potentials / physiology
  • Animals
  • Electromyography / methods*
  • Electrophysiological Phenomena
  • Evoked Potentials / physiology
  • Humans
  • Motor Neurons / physiology*
  • Muscle Fibers, Skeletal / physiology*