A Minimally Invasive Microsensor Specially Designed for Simultaneous Dissolved Oxygen and pH Biofilm Profiling

Sensors (Basel). 2019 Nov 1;19(21):4747. doi: 10.3390/s19214747.

Abstract

A novel sensing device for simultaneous dissolved oxygen (DO) and pH monitoring specially designed for biofilm profiling is presented in this work. This device enabled the recording of instantaneous DO and pH dynamic profiles within biofilms, improving the tools available for the study and the characterization of biological systems. The microsensor consisted of two parallel arrays of microelectrodes. Microelectrodes used for DO sensing were bare gold electrodes, while microelectrodes used for pH sensing were platinum-based electrodes modified using electrodeposited iridium oxide. The device was fabricated with a polyimide (Kapton®) film of 127 µm as a substrate for minimizing the damage caused on the biofilm structure during its insertion. The electrodes were covered with a Nafion® layer to increase sensor stability and repeatability and to avoid electrode surface fouling. DO microelectrodes showed a linear response in the range 0-8 mg L-1, a detection limit of 0.05 mg L-1, and a sensitivity of 2.06 nA L mg-1. pH electrodes showed a linear super-Nernstian response (74.2 ± 0.7 mV/pH unit) in a wide pH range (pH 4-9). The multi-analyte sensor array was validated in a flat plate bioreactor where simultaneous and instantaneous pH and DO profiles within a sulfide oxidizing biofilm were recorded. The electrodes spatial resolution, the monitoring sensitivity, and the minimally invasive features exhibited by the proposed microsensor improved biofilm monitoring performance, enabling the quantification of mass transfer resistances and the assessment of biological activity.

Keywords: DO microsensor; MEMS technology; biofilm monitoring; biofilm profiling; multi-analyte sensor; pH microsensor.

MeSH terms

  • Biofilms*
  • Electrochemistry / instrumentation*
  • Electroplating
  • Equipment Design*
  • Hydrogen-Ion Concentration
  • Microelectrodes
  • Oxygen / analysis*
  • Platinum / chemistry
  • Reproducibility of Results
  • Solubility

Substances

  • Platinum
  • Oxygen