Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2019 Jan-Dec:18:1533033819876897.
doi: 10.1177/1533033819876897.

Factor 10 Expedience of Monthly Linac Quality Assurance via an Ion Chamber Array and Automation Scripts

Affiliations

Factor 10 Expedience of Monthly Linac Quality Assurance via an Ion Chamber Array and Automation Scripts

Lawrie B Skinner et al. Technol Cancer Res Treat. 2019 Jan-Dec.

Abstract

Purpose: While critical for safe and accurate radiotherapy, monthly quality assurance of medical linear accelerators is time-consuming and takes physics resources away from other valuable tasks. The previous methods at our institution required 5 hours to perform the mechanical and dosimetric monthly linear accelerator quality assurance tests. An improved workflow was developed to perform these tests with higher accuracy, with fewer error pathways, in significantly less time.

Methods: A commercial ion chamber array (IC profiler, Sun Nuclear, Melbourne, Florida) is combined with automation scripts to consolidate monthly linear accelerator QA. The array was used to measure output, flatness, symmetry, jaw positions, gated dose constancy, energy constancy, collimator walkout, crosshair centering, and dosimetric leaf gap constancy. Treatment plans were combined with automation scripts that interface with Sun Nuclear's graphical user interface. This workflow was implemented on a standard Varian clinac, with no special adaptations, and can be easily applied to other C-arm linear accelerators.

Results: These methods enable, in 30 minutes, measurement and analysis of 20 of the 26 dosimetric and mechanical monthly tests recommended by TG-142. This method also reduces uncertainties in the measured beam profile constancy, beam energy constancy, field size, and jaw position tests, compared to our previous methods. One drawback is the increased uncertainty associated with output constancy. Output differences between IC profiler and farmer chamber in plastic water measurements over a 6-month period, across 4 machines, were found to have a 0.3% standard deviation for photons and a 0.5% standard deviation for electrons, which is sufficient for verifying output accuracy according to TG-142 guidelines. To minimize error pathways, automation scripts which apply the required settings, as well as check the exported data file integrity were employed.

Conclusions: The equipment, procedure, and scripts used here reduce the time burden of routine quality assurance tests and in most instances improve precision over our previous methods.

Keywords: IC profiler; automation; linac QA; monthly QA; quality assurance.

PubMed Disclaimer

Conflict of interest statement

Declaration of Conflicting Interests: The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Figures

Figure 1.
Figure 1.
A, Ion chamber array layout of IC profiler. B, Schematic monthly QA example setup conditions of detector surface at SDD: 106 cm with varying thickness solid water on top and additional 5 cm solid water underneath. SDD indicates source to detector distance; QA, quality assurance.
Figure 2.
Figure 2.
Top, Comparison of a measured 6 MV inline photon profile for a 25 × 25 cm2 field with 2 different array calibrations: Red lines are the calibration at 34 × 34 cm2, the black lines with circles are the 27 × 27 cm2 calibration. Inset, Close-up of the same data, the measurement with the 34 × 34 cm2 calibration contains significantly more error. Bottom, Several field sizes all with the 34 × 34 cm2 calibration. The arrows indicate the increasing erroneous structure as field size gets further away from the calibration field size. Crossline profiles (not shown) showed the exact same behavior.
Figure 3.
Figure 3.
Flow chart of automation script. Blue boxes indicate automated steps. Black text indicates manual user input or confirmation steps.
Figure 4.
Figure 4.
Central axis dose measurements from the IC profiler. A, Difference of repeated central axis measurements from their average value (same day, no setup changes). This has standard deviations of 0.05% (6E) and 0.1% (6X). B-D, Differences of measured IC profiler output constancy and farmer chamber output over 6 months, on 4 different linacs using (see Methods section for setup detail). Each data point is corrected for machine output as measured by a TG51 protocol in solid water. The results were collected for 6 months.
Figure 5.
Figure 5.
Beam profile measurements compared to the treatment planning system. A) 6MV X-ray field 25 × 25 cm2 field at SSD: 102 and depth 5 cm. B) 12 MeV electron field with 25 × 25 cm2 applicator cone, SSD 105 cm, depth 1.9 cm. The noise in the difference is dominated by the random uncertainty of the eclipse eMC (electron Monte Carlo) calculation. SSD indicates source to detector distance.
Figure 6.
Figure 6.
Collimator walkout measured by IC profiler on 4 different linacs.
Figure 7.
Figure 7.
Dose difference to baseline for sliding leaf gap test. Results from 4 machines of the same type over 1 year.

Similar articles

Cited by

References

    1. Ford EC, Terezakis S, Souranis A, Harris K, Gay H, Mutic S. Quality Control Quantification (QCQ): a tool to measure the value of quality control checks in radiation oncology. Int J Radiat Oncol. 2012;84(3):e263–e269. doi:10.1016/j.ijrobp.2012.04.036. - PubMed
    1. Huq MS, Fraass BA, Dunscombe PB, et al. The report of Task Group 100 of the AAPM: application of risk analysis methods to radiation therapy quality management: TG 100 report. Med Phys. 2016;43(7):4209–4262. doi:10.1118/1.4947547. - PMC - PubMed
    1. Smith K, Balter P, Duhon J, et al. AAPM medical physics practice guideline 8.a.: linear accelerator performance tests. J Appl Clin Med Phys. 2017;18(4):23–39. doi:10.1002/acm2.12080. - PMC - PubMed
    1. Jenkins CH, Naczynski DJ, Yu S-JS, Yang Y, Xing L. Automating quality assurance of digital linear accelerators using a radioluminescent phosphor coated phantom and optical imaging. Phys Med Biol. 2016;61(17):L29–L37. doi:10.1088/0031-9155/61/17/L29. - PMC - PubMed
    1. Clivio A, Vanetti E, Rose S, et al. Evaluation of the machine performance check application for TrueBeam linac. Radiat Oncol. 2015;10(1). doi:10.1186/s13014-015-0381-0. - PMC - PubMed