Corneal Viscous Properties Cannot Be Determined From Air-Puff Applanation
- PMID: 31710375
- DOI: 10.3928/1081597X-20191010-03
Corneal Viscous Properties Cannot Be Determined From Air-Puff Applanation
Abstract
Purpose: To assess whether corneal viscous properties are measureable with air-puff applanation in patients.
Methods: The study had 312 normal eyes, 107 fellow eyes of patients with keratoconus, and 289 keratoconic eyes. The Corvis ST (Oculus Optikgeräte GmbH, Wetzlar, Germany) deformation data for all eyes were analyzed using two models. First, a standard linear solid model (SLM) assumed the cornea was an elastic material only. Second, a two-compartment Kelvin-Voigt model (KVM) assumed the cornea was a visco-elastic material. Corneal stiffness and viscosity were calculated. Further, the deflection amplitude was phase shifted virtually relative to the air-puff applanation force to assess whether the KVM was capable of detecting corneal viscous properties from air-puff applanation. This was similar in concept to measured viscoelastic deformations in other soft tissues. The hysteresis area was also calculated with deformation (cornea and whole globe) and deflection (cornea only) amplitude. The greater the magnitude of the hysteresis area, the greater was the magnitude of corneal viscosity (µc).
Results: Both the SLM and KVM reported similar magnitudes of corneal stiffnesses (correlation coefficient > 0.99). However, for a given model, corneal stiffness was significantly different between normal, fellow, and keratoconic eyes (P = .001). From the KVM, the corneal viscosity was different between groups (P = .001) but was small in magnitude (order of 10-9). The deflection hysteresis area was also small in magnitude (order of 10-6). In contrast, the KVM detected significant corneal viscosity only when the deflection amplitude was virtually phase shifted with respect to the air-puff applanation force.
Conclusions: No significant corneal viscous response was detected in patients who had air-puff applanation. [J Refract Surg. 2019;35(11):730-736.].
Copyright 2019, SLACK Incorporated.
Comment in
-
Can the Corvis ST Estimate Corneal Viscoelasticity?J Refract Surg. 2020 May 1;36(5):346-347. doi: 10.3928/1081597X-20200212-04. J Refract Surg. 2020. PMID: 32396647 No abstract available.
Similar articles
-
Introduction of Two Novel Stiffness Parameters and Interpretation of Air Puff-Induced Biomechanical Deformation Parameters With a Dynamic Scheimpflug Analyzer.J Refract Surg. 2017 Apr 1;33(4):266-273. doi: 10.3928/1081597X-20161221-03. J Refract Surg. 2017. PMID: 28407167
-
Role of Age and Myopia in Simultaneous Assessment of Corneal and Extraocular Tissue Stiffness by Air-Puff Applanation.J Refract Surg. 2016 Jul 1;32(7):486-93. doi: 10.3928/1081597X-20160512-02. J Refract Surg. 2016. PMID: 27400081
-
Variability of Corneal Deformation Response in Normal and Keratoconic Eyes.Optom Vis Sci. 2015 Jul;92(7):e149-53. doi: 10.1097/OPX.0000000000000628. Optom Vis Sci. 2015. PMID: 26002009
-
Biomechanical properties of the keratoconic cornea: a review.Clin Exp Optom. 2015 Jan;98(1):31-8. doi: 10.1111/cxo.12211. Clin Exp Optom. 2015. PMID: 25545947 Review.
-
In vivo human corneal deformation analysis with a Scheimpflug camera, a critical review.J Biophotonics. 2016 May;9(5):464-77. doi: 10.1002/jbio.201500233. Epub 2016 Feb 12. J Biophotonics. 2016. PMID: 26871552 Review.
Cited by
-
The Shape of Corneal Deformation Alters Air Puff-Induced Loading.Front Bioeng Biotechnol. 2022 Mar 30;10:848060. doi: 10.3389/fbioe.2022.848060. eCollection 2022. Front Bioeng Biotechnol. 2022. PMID: 35433651 Free PMC article.
-
Novel Method of Measuring Corneal Viscoelasticity Using the Corvis ST Tonometer.J Clin Med. 2022 Jan 4;11(1):261. doi: 10.3390/jcm11010261. J Clin Med. 2022. PMID: 35012002 Free PMC article.
-
Exploring the Biomechanical Properties of the Human Cornea In Vivo Based on Corvis ST.Front Bioeng Biotechnol. 2021 Nov 17;9:771763. doi: 10.3389/fbioe.2021.771763. eCollection 2021. Front Bioeng Biotechnol. 2021. PMID: 34869287 Free PMC article.
-
A Novel Indentation Assessment to Measure Corneal Biomechanical Properties in Glaucoma and Ocular Hypertension.Transl Vis Sci Technol. 2021 Aug 2;10(9):36. doi: 10.1167/tvst.10.9.36. Transl Vis Sci Technol. 2021. PMID: 34459861 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources