ISCEV extended protocol for the S-cone ERG

Doc Ophthalmol. 2020 Apr;140(2):95-101. doi: 10.1007/s10633-019-09730-6. Epub 2019 Nov 20.

Abstract

The International Society for Clinical Electrophysiology of Vision (ISCEV) standard for full-field electroretinography (ERG) describes a minimum procedure for testing generalized retinal function but encourages more extensive testing. This extended protocol describes a method of assessing the function of the short-wavelength-sensitive cone (S-cone) retinal pathway, using a short-wavelength flash superimposed on a background that saturates the rods and adapts the L/M-cones to elicit a response, known as the S-cone ERG. Stimulus parameters such as the strength and luminance of the flash and background, respectively, and their spectral and temporal characteristics are specified. As a complement to the ISCEV standard, testing the S-cone ERG enables further characterization of light-adapted retinal function and may refine diagnosis of some retinal disorders. Typical applications are described including use in the diagnosis of rod monochromacy and S-cone monochromacy, identification and investigation of cone On-bipolar cell dysfunction and use of the technique to confirm the diagnosis of enhanced S-cone syndrome.

Keywords: Clinical standards; Electroretinogram (ERG); Full-field ERG; International Society for Clinical Electrophysiology of Vision (ISCEV); Retinal dystrophy; Retinopathy; S-cone ERG; Short-wavelength-sensitive cone ERG.

MeSH terms

  • Adaptation, Ocular
  • Calibration / standards
  • Clinical Protocols
  • Electrophysiology / standards*
  • Electroretinography / standards*
  • Humans
  • International Agencies
  • Photic Stimulation
  • Retinal Cone Photoreceptor Cells / physiology*
  • Retinal Dystrophies / physiopathology
  • Rod Opsins / physiology*
  • Societies, Medical / standards*
  • Vision, Ocular

Substances

  • Rod Opsins
  • short-wavelength opsin