Aim: To analyze the accuracy capability of two computer-aided navigation procedures for dental implant placement.
Materials and methods: A total of 40 dental implants were selected, which were randomly distributed into two study groups, namely, group A, consisting of those implants that were placed using a computer-aided static navigation system (n = 20) (guided implant (GI)) and group B, consisting of those implants that were placed using a computer-aided dynamic navigation system (n = 20) (navigation implant (NI)). The placement of the implants from group A was performed using surgical templates that were designed using 3D implant-planning software based on preoperative cone-beam computed tomography (CBCT) and a 3D extraoral surface scan, and the placement of group B implants was planned and performed using the dynamic navigation system. After placing the dental implants, a second CBCT was performed and the degree of accuracy of the planning and placement of the implants was analyzed using therapeutic planning software and Student's t-test.
Results: The paired t-test revealed no statistically significant differences between GI and NI at the coronal (p = 0.6535) and apical (p = 0.9081) levels; however, statistically significant differences were observed between the angular deviations of GI and NI (p = 0.0272).
Conclusion: Both computer-aided static and dynamic navigation procedures allow accurate implant placement.