Molecular detection of the BCR-ABL1 fusion transcripts is necessary for the genetic confirmation of a chronic myeloid leukemia diagnosis and for the risk classification of acute lymphoblastic leukemia. BCR-ABL1 mRNAs are usually identified using a conventional RT-PCR technique according to the BIOMED-1 method. In this study, we evaluated 122 BCR-ABL1-positive samples with the Q-LAMP assay to establish if this technology may represent a valid alternative to the qualitative BIOMED-1 PCR technique usually employed for the detection and the discrimination of the common BCR-ABL1 transcripts (p190 and p210 isoforms). We found a 100% concordance rate between the two methods. Specifically, the p190- and p210-positive samples were amplified by Q-LAMP with a median threshold time (Tt) of 26.70 min (range: 24.45-31.80 min) and 20.26 min (range: 15.25-34.57 min), respectively. A median time of 19.63 was observed in samples displaying both (e13a2/e14a2) p210 isoforms. Moreover, the Q-LAMP assay allowed recognition of the BCR-ABL1 e13a2 and e14a2 isoforms (median Tts 18.48 for e13a2 vs. 26.08 min for e14a2; p < 0.001). Finally, 20 samples harboring rare BCR-ABL1 isoforms (e1a3, e13a3, e14a3, and e19a2) were correctly identified by the Q-LAMP assay. We conclude that the Q-LAMP assay may represent a faster and valid alternative to the qualitative BIOMED-1 RT-PCR for the diagnosis at BCR-ABL1-positive leukemias, especially when samples are analyzed in centers with restricted resources and/or limited technical expertise.
Keywords: BCR-ABL1; Q-LAMP; chronic myeloid leukemia; e13a2; e14a2; rare transcripts.