Observational Study
. 2020 Feb 25;48(2):157-161.
doi: 10.1515/jpm-2019-0357.

Comparison of cord blood lactate measurement by gas analyzer and portable electrochemical devices

Affiliations
Observational Study

Comparison of cord blood lactate measurement by gas analyzer and portable electrochemical devices

Andrey Yurievich Romanov et al. J Perinat Med. .

Abstract

Objective To compare the accuracy of cord blood lactate measurement using gas analyzer and portable devices in order to assess possibility of implementation of these devices in clinical practice. Methods We performed a prospective observational study using 30 umbilical cord samples which were obtained immediately after birth. Portable electrochemical devices Lactate Scout (SensLab GmbH, Leipzig, Germany) and StatStrip Lactate (NOVA Biomedical, Waltham, MA, USA) were used to determine lactate level. A gas analyzer ABL800 FLEX (Radiometer Medical ApS, Brønshøj-Husum, Denmark) was used as a reference. Base excess (BE), pH, partial oxygen (pO2) and carbon dioxide (pCO2) pressure, hemoglobin (ctHb) and bilirubin (ctBl) levels were measured. Results The mean umbilical cord blood lactate level determined by the gas analyzer was 5.85 ± 2.66 mmol/L (ranging from 1.4 mmol/L to 13.4 mmol/L). Lactate level estimated by Lactate Scout was 5.66 ± 2.65 mmol/L and did not significantly differ from the reference method level (P = 0.2547). The mean lactate level determined by StatStrip Lactate was significantly lower than by the gas analyzer - 4.81 ± 2.38 mmol/L (P < 0.0001). Umbilical cord blood pH, BE, pO2 and pCO2, ctHb and ctBl levels did not affect the accuracy of the lactate measurement in absolute units (mmol/L). Conclusion Umbilical cord blood lactate level measured by StatStrip Lactate was lower than estimated by the ABL800 FLEX gas analyzer. This shows the necessity to develop decision-making reference points separately for each device. Umbilical cord blood pH, BE, pO2 and pCO2, ctHb and ctBl levels did not affect the accuracy of measurements by electrochemical portable devices.

Keywords: acid base status; base excess; cord blood lactate; fetal hypoxia; gas analyzer; labor; pH; partial carbon dioxide pressure; partial oxygen pressure.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Perlman JM. Hyperthermia in the delivery: potential impact on neonatal mortality and morbidity. Clin Perinatol 2006;33:55–63.
    1. Perlman JM. Summary proceedings from the neurology group on hypoxic-ischemic encephalopathy. Pediatrics 2006;117:S28–33.
    1. Orsonneau J-L, Fraissinet F, Sébille-Rivain V, Dudouet D, Bigot-Corbel E. Suitability of POC lactate methods for fetal and perinatal lactate testing: considerations for accuracy, specificity and decision making criteria. Clin Chem Lab Med 2013;51:397–404.
    1. Prikhod’ko A, Romanov Ay, Evgrafova A, Shuklina D. Detection of pH and blood lactate levels in the presenting part of the foetus for evaluation of its status during labour. Vopr Ginekol Akusherstva i Perinatol 2017;16:96–9.
    1. East CE, Leader LR, Sheehan P, Henshall NE, Colditz PB, Lau R. Intrapartum fetal scalp lactate sampling for fetal assessment in the presence of a non-reassuring fetal heart rate trace. Cochrane database Syst Rev 2015:CD006174.

Publication types

MeSH terms

LinkOut - more resources