Comparison of Surgical Outcomes Between Standard and Elevated-Rim Highly Cross-Linked Polyethylene Acetabular Liners in Primary Total Hip Arthroplasty With Minimum 15-Year Follow-Up: Single-Center, Retrospective Cohort Study

J Arthroplasty. 2020 May;35(5):1290-1296. doi: 10.1016/j.arth.2019.12.026. Epub 2019 Dec 27.

Abstract

Background: The aims of the present study are to (1) conduct the longest-to-date follow-up wear analysis of the highly cross-linked polyethylene acetabular liners, (2) assess the incidence of femoral and acetabular osteolysis, and (3) compare the surgical outcomes between standard and elevated-rim acetabular liners in primary total hip arthroplasty (THA).

Methods: In this retrospective cohort study, we evaluated 112 primary THAs performed by a single experienced arthroplasty surgeon at our institution between March 2000 and December 2003. Patients were classified based on the type of acetabular liner used: standard or elevated-rim liner. For evaluation of surgical outcomes, the following data were collected: acetabular cup position (anteversion and inclination), wear rate (linear and volumetric), presence of osteolysis, history of reoperation (all-cause and wear-related), complications (deep joint infection, dislocation, and periprosthetic fracture), and Harris hip score at last follow-up.

Results: Linear and volumetric wear rates were 0.028 mm/y (0.000-0.145 mm/y) and 11.641 mm3/y (0.000-70.000 mm3/y) in the standard group and 0.026 mm/y (0.000-0.094mm/y) and 9.706 mm3/y (0.000-33.000 mm3/y) in the elevated-rim group, respectively. These rates were not significantly different between groups. One case of osteolysis was confirmed in the standard group, whereas no osteolysis was observed in the elevated-rim group.

Conclusion: We suggest that elevated-rim highly cross-linked polyethylene acetabular liners might be a good implant option that can be used safely.

Keywords: acetabular liner; elevated-rim; highly cross-linked polyethylene; liner wear; total hip arthroplasty.

MeSH terms

  • Arthroplasty, Replacement, Hip* / adverse effects
  • Follow-Up Studies
  • Hip Prosthesis*
  • Humans
  • Osteolysis* / epidemiology
  • Osteolysis* / etiology
  • Polyethylene
  • Prosthesis Design
  • Prosthesis Failure
  • Reference Standards
  • Retrospective Studies
  • Treatment Outcome

Substances

  • Polyethylene