Impact of turbulent blood flow in the aortic root on de novo aortic insufficiency during continuous-flow left ventricular-assist device support
- PMID: 32080864
- DOI: 10.1111/aor.13671
Impact of turbulent blood flow in the aortic root on de novo aortic insufficiency during continuous-flow left ventricular-assist device support
Abstract
Severe aortic insufficiency (AI) after implantation of continuous-flow left ventricular-assist device (LVAD) affects device performance and outcomes. However, the mechanism for the occurrence and progression of AI has not been elucidated. We investigated the impact of nonphysiological retrograde blood flow in the aortic root on AI after LVAD implantation. Blood flow pattern was analyzed in patients with and without AI (n = 3 each) who underwent LVAD implantation, by computational fluid dynamics with patient-specific geometries, which were reproduced using electrocardiogram-gated 320-slice computed tomographic images. The total volume of retrograde blood flow during one cardiac cycle (716 ± 88 mL) was higher and the volume of slow blood flow (<0.1 cm/s) (0.16 ± 0.04 cm3 ) was lower in patients with AI than in those without AI (360 ± 111 mL, P = .0495, and 0.49 ± 0.08 cm3 , P = .0495, respectively). No significant difference in wall shear stress on the aortic valve was observed between the groups. Patients with AI had a perpendicular anastomosis at the distal ascending aorta and the simulation in the modified anastomosis model of patients with AI showed that the retrograde blood flow pattern depended on the angle and position of anastomosis. Computational fluid dynamics revealed strong retrograde blood flow in the ascending aorta and aortic root in patients with AI after LVAD implantation. The angle and position of LVAD outflow anastomosis might impact retrograde blood flow and de novo AI after LVAD implantation.
Keywords: blood flow analysis; computational fluid dynamics; de novo aortic insufficiency; left ventricular assist device.
© 2020 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.
Similar articles
-
Computational fluid dynamic study of hemodynamic effects on aortic root blood flow of systematically varied left ventricular assist device graft anastomosis design.J Thorac Cardiovasc Surg. 2015 Sep;150(3):696-704. doi: 10.1016/j.jtcvs.2015.05.034. Epub 2015 May 15. J Thorac Cardiovasc Surg. 2015. PMID: 26092505
-
Use of patient-specific computational models for optimization of aortic insufficiency after implantation of left ventricular assist device.J Thorac Cardiovasc Surg. 2021 Nov;162(5):1556-1563. doi: 10.1016/j.jtcvs.2020.04.164. Epub 2020 May 15. J Thorac Cardiovasc Surg. 2021. PMID: 32653292 Free PMC article.
-
Computational fluid dynamics in patients with continuous-flow left ventricular assist device support show hemodynamic alterations in the ascending aorta.J Thorac Cardiovasc Surg. 2014 Apr;147(4):1326-1333.e1. doi: 10.1016/j.jtcvs.2013.09.069. Epub 2013 Dec 15. J Thorac Cardiovasc Surg. 2014. PMID: 24345553
-
De novo aortic insufficiency during long-term support on a left ventricular assist device: a systematic review and meta-analysis.ASAIO J. 2014 Mar-Apr;60(2):183-8. doi: 10.1097/MAT.0000000000000042. ASAIO J. 2014. PMID: 24399060 Review.
-
Preoperative Prediction of Aortic Insufficiency During Ventricular Assist Device Treatment.Int Heart J. 2016;57(1):3-10. doi: 10.1536/ihj.15-250. Epub 2016 Jan 6. Int Heart J. 2016. PMID: 26742702 Review.
Cited by
-
The Impact of Left Ventricular Assist Device Outflow Graft Positioning on Aortic Hemodynamics: Improving Flow Dynamics to Mitigate Aortic Insufficiency.Biomimetics (Basel). 2023 Oct 1;8(6):465. doi: 10.3390/biomimetics8060465. Biomimetics (Basel). 2023. PMID: 37887596 Free PMC article.
-
Preoperative higher right ventricular stroke work index increases the risk of de novo aortic insufficiency after continuous-flow left ventricular assist device implantation.J Artif Organs. 2023 Jul 19. doi: 10.1007/s10047-023-01411-1. Online ahead of print. J Artif Organs. 2023. PMID: 37468735
-
Aortic valve disorders and left ventricular assist devices.Front Cardiovasc Med. 2023 Feb 23;10:1098348. doi: 10.3389/fcvm.2023.1098348. eCollection 2023. Front Cardiovasc Med. 2023. PMID: 36910539 Free PMC article. Review.
-
Concomitant or late aortic valve intervention and its efficacy for aortic insufficiency associated with continuous-flow left ventricular assist device implantation.Front Cardiovasc Med. 2022 Nov 15;9:1029984. doi: 10.3389/fcvm.2022.1029984. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 36457799 Free PMC article. Review.
-
Prediction, prevention, and management of right ventricular failure after left ventricular assist device implantation: A comprehensive review.Front Cardiovasc Med. 2022 Nov 3;9:1040251. doi: 10.3389/fcvm.2022.1040251. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 36407460 Free PMC article. Review.
References
REFERENCES
-
- Slaughter MS, Rogers JG, Milano CA, Russell SD, Conte JV, Feldman D, et al. Advanced heart failure treated with continuous-flow left ventricular assist device. New England J Med. 2009;361(23):2241-51.
-
- Toda K, Fujita T, Domae K, Shimahara Y, Kobayashi J, Nakatani T. Late aortic insufficiency related to poor prognosis during left ventricular assist device support. Ann Thoracic Surg. 2011;92(3):929-34.
-
- Aggarwal A, Raghuvir R, Eryazici P, Macaluso G, Sharma P, Blair C, et al. The development of aortic insufficiency in continuous-flow left ventricular assist device-supported patients. Ann Thoracic Surg. 2013;95(2):493-8.
-
- Hata H, Fujita T, Ishibashi-Ueda H, Nakatani T, Kobayashi J. Pathological analysis of the aortic valve after long-term left ventricular assist device support. Eur J Cardio-Thoracic Surg. 2014;46(2):193-7.
-
- Yoshida S, Fukushima S, Miyagawa S, Toda K, Sawa Y. Visualization of vortex flow and shear stress in the aortic root during left ventricular assist device support. J Thoracic Cardiovasc Surg. 2017;154(3):877-8.e871.
MeSH terms
LinkOut - more resources
Full Text Sources