Design of a Fully Integrated Inductive Coupling System: A Discrete Approach Towards Sensing Ventricular Pressure

Sensors (Basel). 2020 Mar 10;20(5):1525. doi: 10.3390/s20051525.

Abstract

In this paper, an alternative strategy for the design of a bidirectional inductive power transfer (IPT) module, intended for the continuous monitoring of cardiac pressure, is presented. This new integrated implantable medical device (IMD) was designed including a precise ventricular pressure sensor, where the available implanting room is restricted to a 1.8 × 1.8 cm2 area. This work considers a robust magnetic coupling between an external reading coil and the implantable module: a three-dimensional inductor and a touch mode capacitive pressure sensor (TMCPS) set. In this approach, the coupling modules were modelled as RCL circuits tuned at a 13.56 MHz frequency. The analytical design was validated by means of Comsol Multiphysics, CoventorWare, and ANSYS HFSS software tools. A power transmission efficiency (PTE) of 94% was achieved through a 3.5 cm-thick biological tissue, based on high magnitudes for the inductance (L) and quality factor (Q) components. A specific absorption rate (SAR) of less than 1.6 W/Kg was attained, which suggests that this IPT system can be implemented in a safe way, according to IEEE C95.1 safety guidelines. The set of inductor and capacitor integrated arrays were designed over a very thin polyimide film, where the 3D coil was 18 mm in diameter and approximately 50% reduced in size, considering any conventional counterpart. Finally, this new approach for the IMD was under development using low-cost thin film manufacturing technologies for flexible electronics. Meanwhile, as an alternative test, this novel system was fabricated using a discrete printed circuit board (PCB) approach, where preliminary electromagnetic characterization demonstrates the viability of this bidirectional IPT design.

Keywords: BioMEMS; flexible electronics; integrated coupling system; ventricular pressure sensor; wireless power transfer, implantable medical device.

MeSH terms

  • Adipose Tissue / pathology
  • Electric Power Supplies
  • Electronics
  • Electrophysiology / instrumentation*
  • Electrophysiology / methods
  • Equipment Design
  • Heart Ventricles*
  • Humans
  • Magnetics / instrumentation
  • Muscles / pathology
  • Patient Safety
  • Prostheses and Implants
  • Signal Processing, Computer-Assisted
  • Skin / pathology
  • Ventricular Pressure*
  • Wireless Technology / instrumentation